Swann AMU-II Oxytrace Operator's Manual page 66

Table of Contents

Advertisement

AMU-II Oxytrace
Program List and Explanations
5.3.2.32.4
5.3.2.1
5.3.2.24
5.3.2.24
5.3.2.340
5.3.2.44
5.3.2.54
5.3.2.6
5.3.2.7
66
Control Parameters:
Range for each Parameter same as
Function = Timer
The relay will be activated repetitively depending on the programmed
time scheme.
Mode: Operating mode (interval, daily, weekly)
Interval
Interval: The interval can be programmed within a range
of 1–1'440 min.
Run Time: Enter the time the relay stays active.
Range: 5–32'400 sec.
Delay: during run time plus the delay time the signal and control out-
puts are held in the operating mode programmed below.
Range: 0–6'000 Sec.
Signal Outputs: Select operating mode of the signal output:
Signal outputs continue to issue the measured value.
Cont.:
Signal outputs hold the last valid measured value.
Hold:
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Signal outputs are switched off (set to 0 or 4 mA).
Off:
Errors, except fatal errors, are not issued.
Output/Control: Select operating mode of the controller output:
Cont.:
Controller continues normally.
Controller continues based on the last valid value.
Hold:
Controller is switched off.
Off:
5.2.1.43, p. 60
A-96.150.541 / 111223