AMU-II Oxytrace
Error List
Error
E013
Case Temp. high
E014
Case Temp. low
E017
Control Timeout
E019
Quality Assurance
E024
Input active
E026
IC LM75
E030
EEProm Frontend
E031
Calibration Recout
E032
Wrong Frontend
E033
Power-on
E034
Power-down
E065
Electrolyte depleted
46
Description
Corrective action
– check case/environment temperature
– check programmed value,
see
5.3.1.5.1, p. 63
– check case/environment temperature
– check programmed value,
see
5.3.1.5.2, p. 63
– check control device or programming
in Installation, Relay contact, Relay 1/2
see
5.3.2 and 5.3.3, p. 63
– Perform QA procedure using a
reference instrument, e.g. AMI
Inspector
– See If Fault Yes is programmed in
Menu see
5.3.4, p. 68
– call service
– call service
– call service
– call service
– none, normal status
– none, normal status
– Refill electrolyte, see
exchange, p. 32
Electrolyte
A-96.150.541 / 111223