AMU-II Oxytrace
Program List and Explanations
5.3.4
5.3.4.1
5.3.4.2
5.3.4.3
5.3.4.4
5.3.4.5
5.4 Miscellaneous
5.4.1
68
Input: The functions of the relays and signal outputs can be defined
depending on the position of the input contact, i.e. no function,
closed or open.
Active: Define when the input should be active:
Input is never active.
No:
Input is active if the input relay is closed
When closed
Input is active if the input relay is open
When open:
Signal Outputs: Select the operation mode of the signal outputs
when the relay is active:
Signal outputs continue to issue the measured
Cont.:
value.
Signal outputs issue the last valid measured value.
Hold:
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Output/Control: (relay or signal output):
Controller continues normally.
Cont.:
Controller continues on the last valid value.
Hold:
Controller is switched off.
Off:
Fault:
No message is issued in pending error list and the
No:
alarm relay does not close when input is active.
Message E024 is stored in the message list.
Message E024 is issued and stored in the mes-
Yes:
sage list. The Alarm relay closes when input is
active.
Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6'000 Sec
Language: Set the desired language.
Available settings: German, English, French , Spanish, Chinese
A-96.150.541 / 111223