Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

10 Door Interlocks

The door interlocks is wired to the C2 cabinet.
194
IEC 60601–2–43 requires not to install door interlocks. It is the responsibility of the
field service to check that this requirement is not in contradiction with local
regulation. In case of conflict, follow local regulation. No other measures employed
for radiation protection should cause the interruption of irradiation and any other
disturbance of a procedure in progress.
Direction 5160944-13-1EN, Revision 2