NOTE:
(3): In use temperature limits specify the range where the system shall work. Operating
outside these limits could occur severe performance and reliability issues.
Recommended temperature limits specify the range where it is recommended to adjust
air conditioning control in order to warranty current operations inside the in use range.
Relative Humidity and Temperature: Refer to
temperature requirements for components not specified in
component Pre–Installation Manual listed in
Compatibility.
Altitude and Atmospheric Pressure: Refer to
pressure requirements for components not specified in
component Pre–Installation Manual listed in
Compatibility.
Table 3-2:
INSTALLATION ROOM
OF PRODUCT OR COM‐
PONENT
Examination room
Technical room (See No‐
tice below
Control room
Chapter 3 Physical Requirements of Room
ALTITUDE (meters)
IN-USE
MIN
MAX
0
3000
0
3000
0
3000
The chiller is able to dissipate maximum continuously power at 3000 meters in an
ambient temperature up to 20°C, for this altitude the technical room temperature
shall not exceed 20°C.
In some cases condensation occurs and water drops from outlets and pipes of the
air conditioner in the technical room.
Therefore, it is critical to install the cabinets where there is no risk of flood from the
air conditioner.
Ensure the air outlet positions in the exam room is not in area closed to monitor
suspension carriage/rail.
Table
3-1. To obtain relative humidity and
Chapter 2, Section 3, Basic Innova System
Table
3-2. To obtain altitude and atmospheric
Table
Chapter 2, Section 3, Basic Innova System
STORAGE / TRANS‐
PORT
MIN
MAX
MIN
-30
3000
70
-30
3000
70
0
3000
N/A
Direction 5160944-13-1EN, Revision 2
Table
3-1, refer to the appropriate
3-2, refer to the appropriate
ATMOSPHERIC PRESSURE (kPa)
IN-USE
STORAGE / TRANSPORT
MAX
MIN
106
70
106
50
N/A
N/A
MAX
106
106
N/A