Electrical Requirements - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
NOTE:
NOTE:
Ensure sufficient cable length on the Procedures Room end of the cables according to the
table below
* Consider the 12' length of the PIMr cable, if PIMr exists.
** Pimette has 3 m cable.
NOTE:
The incorrect cable length might cause reliability issues during the operation !
Make sure, that the out of conduit parts of the cables with the connectors are properly protected
during the rest of the installation

6.4 Electrical Requirements

The power for the IVUS hardware components delivered with the product (PC, monitor, Video
Switch and printer) is supplied through the IVUS Safety Isolation Transformer.
The customer is in charge to provide a wall outlet to supply the Safety Isolation Transformer. This
outlet shall be located in the control room suitable to the IVUS components positioning. The
maximum rated power of the Safety Isolation Transformer is 600 VA, the maximum continuous
system power consumption for s5i and s5iGE IVUS systems is 400 VA.
The s5i and s5iGE IVUS systems can be configured to work on 100, 120, or 230 VAC. The
difference between the supported configurations is limited switching the fuse values. See table
below :
Frequency: 50/60 Hertz
The country specific power cords are pre-selectable in the GTC.
Chapter 6 Electrical Connections
You should consider running the ECG cable through the conduit in case, the
customer requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future
expansion options.
Same applies to the 2nd Ethernet cable, if a later extension of the IVUS remote
controls expected.
Cable
IVUS PIM cable *
FFR cables for Pimette **
Control Station cable
Joystick cable
ECG cable
Line Voltage
100 VAC
120 VAC
230 VAC
Direction 5160944-13-1EN, Revision 2
Free length from the conduit exit
2.4 m (8 ')
0m (0')
2.1 m (7 ')
2.7 m (9 ')
Depends on the ECG location
Fuse value
8 A
6.3 A
3.15 A
189