Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

2.1 Room layout and room min length

Table 4-1: Exam Room dimensions
Recommended:
Minimum:
Note (1): With a 3 monitor frame suspension, take care with the ceiling height not more 3.05 m (10 ft)
Illustration 4-2: Omega V or Innova IQ Table (Angio, Cardio, Neuro) with header extender
90
Length
32 ft 0 in (9.75m)
For minimum exam room lengths, see different table configu‐
rations in
Illustration
4-2,
Illustration 4-3
Direction 5160944-13-1EN, Revision 2
20 ft 0 in ( 6.00 m)
14 ft 5 in (4.40 m)
and
Illustration 4-4
Width
Ceiling Height
10 ft 0 in (3.05 m)
9 ft 0 in (2.74 m)