7 X-Ray Lamp Room Distribution
The fitter shall install a lamp in:
Exam room,
Control room.
So that they are connected to the equipment through connexion terminals.
Innova LC provides a room distribution from C2 cabinet as follows:
X–ray–on–line (red) drive by low voltage relay 24 V. Acting during X–Ray exposures in Control
Room.
X–ray–on–line (red) drive by low voltage relay 24 V.
In the Exam Room, visible from any place in the room.
Acting during X–ray exposures.
room light (limited to 24 Vac maxi, wires sized to 10 Amps maxi), controls an external relay
which applies power to the room lamp.
In the control room, an additional X-Ray on light must be installed if the console (VCIM) indicator
cannot be perceived by all the persons in the control room.
Chapter 6 Electrical Connections
THE X-RAY ON LAMP MUST BE INSTALLED IN THE EXAM ROOM TO
CONFORM TO INTERVENTIONAL STANDARD IEC/EN 60601-2-43.
SIGNAL INDICATING THE X-RAY ON SHALL BE PERCEPTIBLE BY THE
OPERATOR IN ALL THE LOCATIONS DEFINED FOR THE PERSONNEL WHO
MAY RECEIVE SCATTERED RADIATION
Direction 5160944-13-1EN, Revision 2
191