2285646 - Through Bolts Without Insert - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
2.3.3.2 2285646 – Through bolts without insert
Table 5-9:
Item
156
Name
Part #
Bolt, Hex
2296892
Washer, Flat 99142204
Plate
2290941
Nut, Hex
99141607
Grout assy
2285055
Dowel
2290937
Description
Through Bolt; M20 - 500-400
12 bolts for LC Positioner and
4 bolts for Table
Special Flat Washer for Through
Bolts; one for each bolt
Special Steel Spacer Plate; 4 in. x
4 in. (102 mm x 102 mm); one for
each bolt
Hex Nut M20 STL galvanized two
for each bolt
- 10 kg Powdered Mortar Ardex
K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
Wood Dowel; 24 mm diameter
Direction 5160944-13-1EN, Revision 2
Quan.
16 pc
Through Bolt mounting
method hardware
16 pc
same as above
16 pc
same as above
32 pc
same as above
1 kit
Used in constructing LC Po‐
sitioner grout dam
12 pc
2 Mounting Requirements
Notes

Advertisement

Table of Contents
loading

This manual is also suitable for:

Innova 3100Innova 3100-iqInnova 4100Innova 4100-iq

Table of Contents