11 Emergency

11.1 Partial UPS Fluoro (option)

A Fluoro UPS (20 kVA) has been designed for the Innova system. This partial UPS lets the
customer complete an exam in fluoro mode in case of a power failure. The autonomy provided by
this UPS is 5 minutes of fluoro every 24 hours.

11.2 Full UPS Record (option)

An UPS sized for 150 kVA is required to supply the system in record mode. Such an UPS will
provide to the customer about 10 minutes of autonomy. Contact your local GEMS representative
for more information and/or public power distribution analysis.
NOTE:
An UPS is inserted inside C1 Cabinet and the DL is not affected by microcuts with a
duration of less than 10µs.
C1 UPS accepts a power cut without a hardware risk.
The UPS inserted inside C1 Cabinet is a part of the system and it is not permitted to
replace it by another locally supplied unit.
UPS autonomy is 1 minute before shutdown. Time before shut down is 1 minute without
risk of loosing stored image.
Time to completely charge the battery of UPS is 48 hours.