Electromagnetic Immunity - GE Innova 2100-IQ Preinstallation Manual

Cardiovascular imaging system
Hide thumbs Also See for Innova 2100-IQ:
Table of Contents

Advertisement

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare

2.4.3 Electromagnetic Immunity

2.4.3.1 Electromagnetic Immunity
The Innova system is suitable for use in the specified electromagnetic environment. The purchaser
or user of the Innova system should assure that it is used in an electromagnetic environment as
described below:
Table 3-10:
Immunity Test
Electrostatic dis‐
charge (ESD)
IEC 61000–4–2
Electrical fast transi‐
ent/burst
IEC 61000–4–4
Surge
IEC 61000–4–5
Voltage dips, short in‐
terruptions and volt‐
age variations on
power supply input
lines
IEC 61000–4–11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000–4–8
Note: These are guidelines. Actual conditions may vary.
Chapter 3 Physical Requirements of Room
IEC 60601–1–2 Test
Compliance Level
Level
6 kV contact
6 kV contact
8 kV air
8 kV air
2 kV for power supply
2 kV for power supply
lines
lines
1 kV for input/output
1 kV for input/output
lines
lines
1 kV line(s) to lines(s)
1 kV line(s) to lines(s)
2 kV line(s) to earth
2 kV line(s) to earth
0 % U n for 5 sec
0 % U n for 5 sec
3 A/m
3 A/m
1A/m
Direction 5160944-13-1EN, Revision 2
Electromagnetic Environment
Floors are wood, concrete, or ceramic tile, or floors are covered
with synthetic material and the relative humidity is at least 30 per‐
cent.
Mains power quality is that of a typical commercial and/or hospital
environment
Mains power quality is that of a typical commercial and/or hospital
environment.
Mains power quality is that of a typical commercial and/or hospital
environment. If the user of the Innova system requires continued
operation during power mains interruptions, it is recommended
that the Innova system be powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields are at levels characteristic of a
typical location in a typical commercial and/or hospital environ‐
ment.
At that disturbance amplitude level the monitor image might
present some slight flicker. For image quality improvement, the
monitor might be removed from the low frequency magnetic field
source vicinity.
At that disturbance amplitude level the system is fully operational.