NOTE:
For more details on Floor plate or Baseplates, refer
Holes location in concrete floor in
NOTE:
With any kind of fixation methods (Bolts M20, Mechanical anchors or Chemical
anchors), the number of holes used mandatory is:
we can have only 2 consecutive holes omitted.
NOTE:
In case of floor plate configuration, the gantry baseplate shall be mounted onto the floor
plate, but the table baseplate shall not be mounted on the floor plate.
Pull out efforts and recommendations about chemical anchors not provided by GE.
The following table provides the recommended chemical anchors for Omega/LC baseplates and
for the floor plate ordered locally that they could be used instead of bolts provided by GE.
Table 5-4: Chemical anchors Pull out efforts and recommendations
Mark
Pull out effort
Number of holes in the
plate
Recommended chemi‐
cal anchors example 1
Threaded rod
Hole diameter in the
floor
Hole depth in the floor
Minimum floor thickness
Max Tightening Torque
Chapter 5 Physical Characteristics
The Table baseplate or the System baseplate are mandatory to install the table
(patient support).
The Table must never be installed on grade.
Gantry baseplate : 12 max and 8 min holes used are acceptable
Table baseplate : 10 max and 8 min holes used are acceptable
Floor baseplate : 24 max and 12 min holes used are acceptable
Gantry baseplate
I on
Illustration 5-32
736 daN per bolt if 12
used and 1992 daN per
bolt if 8 used
12 max (8 min mandato‐
ry)
Supplier HILTIHVU ad‐
hesive capsule + HAS
Anchor rod
M16 A4-70 / 333 131 5/8 M20 A4-70 / 333 135 3/4 M20 A4-70 / 333 135 3/4 M16 A4-70 / 333 131 5/8
18 mm
(11/16 in)
125 mm
(5 in)
180 mm
(7 in)
80 N.m
(59 ft-lb)
Chapter 3, Structural
Table baseplate Omega
J on
Illustration 5-32
A on
1120 daN per bolt if 10
4432 daN per bolt with 4
used and 2000 daN per
bolt if 8 used
10 max (8 min mandatory)
Supplier HILTIHVU adhe‐
Supplier HILTIHVU adhe‐
sive capsule + HAS An‐
sive capsule + HAS An‐
chor rod
chor rod
24 mm
(7/8 in)
170 mm
(6-5/8 in)
220 mm
(8-1/2 in)
150 N.m
(110 ft-lb)
Direction 5160944-13-1EN, Revision 2
toIllustration 5-33
Requirements.
Table Omega
Floor plate (to be ordered
Illustration 5-33
K on
Illustration 5-32
272 daN per bolt if 24 used
bolts
and 2008 daN per bolt if 12
4 mandatory
24 max (12 min mandatory)
Supplier HILTIHVU adhe‐
sive capsule + HAS Anchor
rod
24 mm
(7/8 in)
170 mm
(6-5/8 in)
220 mm
(8-1/2 in)
150 N.m
(110 ft-lb)
and illustration
locally)
used
18 mm
(11/16) in
125 mm
(5 in)
180 mm
(7 in)
80 N.m
(59 ft-lb)