To Load A File; Table 23.4 Waveform Data Header - Rigol DHO900 Series User Manual

Digital oscilloscope
Hide thumbs Also See for DHO900 Series:
Table of Contents

Advertisement

Store and Load
Model
Channel Name
3. Waveform Data Header
A waveform may have multiple data sets. Each waveform data set has a waveform
data header. The waveform data header consists of information about the
waveform data set. The header is stored before the data set.

Table 23.4 Waveform Data Header

Header Size
Buffer Type
Bytes Per Point
Buffer Size
23.3

To Load a File

In the storage setting menu, click or tap the Load tab to switch to the load menu. In
this menu, you can load the local file to the instrument.
DHO900 User Guide
242
A 24-byte character array in the format of MODEL#:SERIAL#,
indicating the oscilloscope's model and serial number.
A 16-byte character array that contains the label assigned to
the waveform.
A 4-byte integer, indicating the number of bytes in the
waveform data header.
A 2-byte integer, indicating the type of the waveform data
stored in the file.
- 0 = Unknown
- 1 = Normal 32-bit float data
- 2 = Maximum float data
- 3 = Minimum float data
- 4 = Not Used
- 5 = Digital unsigned 8-bit character data (for digital
channels)
A 2-byte short integer, indicating the number of bytes per
data point.
An 8-byte long integer, indicating the number of bytes of the
current channel waveform data.