Measurement Settings; Figure 11.2 Frequency Counter Setting Menu - Rigol DHO900 Series User Manual

Digital oscilloscope
Hide thumbs Also See for DHO900 Series:
Table of Contents

Advertisement

You can click or tap the "Counter" label in the "Result" bar and select Reset Stat.,
Setting, or Remove in the displayed window.
11.2.1

Measurement Settings

After the frequency counter is enabled, click or tap the "Counter" label in the "Result"
bar at the right section of the screen and a window is displayed. Click or tap Setting
in the window to enter the counter setting menu, as shown in the figure below.
Select the Measurement Source
Click or tap the drop-down button of Source to select the desired source. Analog
channels (CH1-CH4) and digital Channels (D0-D15) can be selected as the source of
the frequency counter.
NOTE
Only when the digital channel probe is connected, can you select digital channels D0-D15 as
the source.
Set Resolution
For Period and Frequency measurements, you need to set the readout resolution.
Click or tap the input field of Resolution to set the resolution by using the pop-up
numeric keypad. You can also use the corresponding multipurpose knob to set the
value. The range of resolution is from 3 bits to 6 bits. By default, it is 4 bits.
The greater the resolution, the longer the gate time. In this way, the measurement
time will be longer.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Digital Voltmeter (DVM) and Frequency Counter

Figure 11.2 Frequency Counter Setting Menu

DHO900 User Guide