Selfcal; Option List; Quick Action Settings - Rigol DHO900 Series User Manual

Digital oscilloscope
Hide thumbs Also See for DHO900 Series:
Table of Contents

Advertisement

24.6

SelfCal

The self-calibration program can quickly make the oscilloscope to work in an optimal
state to get the precise measurement results. You can perform self-calibration at any
time, especially when the changes of the ambient temperature reach or exceed 5℃.
Make sure that the oscilloscope has been warmed up or operating for more than 30
minutes before the self-calibration.
In "Utility" menu, click or tap SelfCal, the self-calibration menu is as shown below.
Click or tap Start, and then the oscilloscope will start to execute the self-
calibration program.
After starting the self-calibration program, click or tap Exit to cancel self-
calibration operation at any time.
Click or tap Close to close the self-calibration information window.
24.7

Option List

In the "Utility" menu, click or tap Options to view all the options of the instrument.
24.8

Quick Action Settings

In Utility menu, click or tap Quick to enter the menu to configure the quick action
key on the front panel.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Figure 24.1 Self-calibration Menu
System Utility Function Setting
DHO900 User Guide
253