3. Use the probe to connect the input terminal of CH1 and the "Compensation Signal
Output Terminal" of the probe, as shown in
4. Set the probe ratio based on the attenuation of the probe, and then click or tap
> Auto.
5. Observe the waveform on the display. In normal condition, you should see a
square waveform similar to the waveform shown in the figure below.
6. Test the other channels in the same way. If you see the waveform, but the square
wave is not shaped correctly as shown above, perform the procedure described in
Probe Compensation
WARNING
To avoid electric shock when using the probe, please make sure that the insulated wire of
the probe is in good condition. Do not touch the metallic part of the probe when the
probe is connected to high voltage source.
DHO900 User Guide
16

Figure 4.7 Using the Compensation Signal

Figure 4.8 Square Waveform Signal

. If you do not see the waveform, perform those steps again.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Figure 4.7
.
Compensation Signal
Output Terminal
Ground Terminal