Measurement Settings

In the Measure menu, click or tap the Setting button to enter the measurement
setting menu.

Figure 10.7 Measurement Settings Menu

Indicator
In the Measure setting menu, click or tap the Indicator on/off switch to enable or
disable the indicator.
If enabled, one or more cursors will be displayed on the screen. Before enabling the
indicator, you need to enable at least one auto measurement parameter and the
number of cursors will change with the measurement parameter enabled.
TIP
When no measurement parameter is selected or there is no source input, the indicator is not
available. The indicator changes when the waveform is expanded or compressed horizontally.
Measurement Threshold
First, select % or Abs as the display type.
Click or tap the drop-down button of Source to select the desired channel
(CH1~CH4 or Math1~Math4).
Click or tap the input field of Upper and use the pop-up numeric keypad to set
the upper limit of the measurement or use the corresponding multipurpose