RIGOL
The unit is V*s. The area of the waveform above the zero reference (namely the
vertical offset) is positive, and the area of the waveform below the zero
reference is negative. The area measured is the algebraic sum of the whole
period area.

Measurement Settings

1. Measurement Category
Press Measure → Add → Category to select "Horizontal", "Vertical", or
"Other". Meanwhile, the specific measurement parameters are displayed on the
screen, as shown in the figure below. In the measurement category area, you
can also rotate the multifunction knob
"Other". Then press down the knob to switch to the desired measurement
category.
"Horizontal" measurement parameters: Period, Freq, RiseTime, FallTime,
+Width, -Width, +Duty, -Duty, +Pulses, -Pulses, +Edges, -Edges, Tvmax,
Tvmin, +SlewRate, and –SlewRate.
"Vertical" measurement parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp,
Vmid, Vlower, Vavg, VRMS, Per.VRMS, Overshot, Preshoot, Area, Per.Area,
and Variance.
"Other" measurement parameters: Delay(A →B ), Delay(A →B ),
Delay(A →B ), Delay(A →B ), Phase(A →B ), Phase(A →B ),
Phase(A →B ), and Phase(A →B ).
Measurement Category Area Measurement Parameter Area
In the measurement parameter area, rotate the multifunction knob
a measurement item and then press down the knob to select it. Then the
measurement results of the item are displayed at the bottom of the screen
(measurement results of up to 10 items can be displayed at most). When you
add a new measurement item, its measurement result will be displayed as the
first, with the original items moving right in sequence. If the number of
measurement items added exceeds 10, then when you add one item, the
6-38