USING THE SUPPLY FUNCTIONS

This option lets you modify specific components and parameters of your supply. The
menu options are:

Setting the AE Bus Timeout

SUPPLY → AE BUS
To set your AE Bus timeout time:
Note: We recommend you use the default value of 200 ms.

Setting the Supply Parameters

SUPPLY → DEFAULTS
To set all your supply parameters to the defaults and then store them in non-
volatile memory:

Setting Pulse Parameters

SUPPLY → PULSE
5702269-C
Page 73 of 204
5. To display the software version of the DSP software, press DSP.
6. To display the software version of the logic, press LOGIC.
• AEBUS:
• DEFAULTS:
• PULSE:
• PANEL:
• PGM_SRC:
1. Press SUPPLY → AEBUS.
2. Rotate the MODIFY knob to adjust the AE Bus time out value (in 10 ms
increments).
3. Press ENTER to accept and exit.
1. Press SUPPLY → DEFAULTS.
2. Rotate the MODIFY knob to display YES or NO.
3. Press ENTER to accept and exit.
I/O Communication and Control
Pinnacle™ Plus+ 10 kW
4-29
Ex. 1024 Page 73