Figure 8.12 Duration Trigger Setting Menu - Rigol DHO924S User Manual

Digital oscilloscope
Hide thumbs Also See for DHO924S:
Table of Contents

Advertisement

Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
1.888.610.7664

Figure 8.12 Duration Trigger Setting Menu

After a trigger type is selected, the current trigger setting information (including the
trigger type, source, and level) is displayed in the trigger information label at the top
of the screen, as shown in the figure below. The information will change based on the
trigger settings.
Source Selection
Click or tap the Source drop-down button to select CH1-CH4 or D0-D15. For details,
refer to
Trigger Source
. The selected trigger source is indicated in the trigger
information label at the top of the screen.
Only when you select the channel that has signal inputs as the trigger source, can you
obtain a stable trigger.
Pattern Setting
The following three patterns are available:
1: sets the pattern of the channel selected to "1", i.e. the voltage level is higher
than the trigger level of the channel.
0: sets the pattern of the channel selected to "0", i.e. the voltage level is lower
than the trigger level of the channel.
X: sets the pattern of the channel selected to "X", i.e. this channel is not used as a
part of the pattern. When all channels in the pattern are set to "X", the
oscilloscope will not trigger.
The Left/Right arrow key indicates moving left/right to switch the channel pattern.
"All" indicates all bits. Select a pattern for a channel and then click or tap All. The
www.calcert.com
15
10
20
5
25
0
30
DHO900 User Guide