Bode Plot (For Dho914S And Dho924S Only); Figure 15.1 "Bode Plots" Setting Menu; Figure 15.2 "Bode Plots" Setting Menu-Simplified Mode - Rigol DHO924S User Manual

Digital oscilloscope
Hide thumbs Also See for DHO924S:
Table of Contents

Advertisement

15
DHO900 User Guide
1.888.610.7664
178

Bode Plot (for DHO914S and DHO924S only)

Bode plot is a way of graphically displaying the frequency response of a system. In
the switch power supply and operational amplifier's circuit feedback network, the
Bode plot provides the curves displaying the variation of gain and phase with the
frequency for a loop analysis. Through the analysis on the system's gain and phase
margins, you can determine the stability of the system.
With the built-in signal generator module, the DHO914S and DHO924S series
generate the sweep signal of a specified frequency range and outputs to the switch
power supply circuit under test. Then, the oscilloscope draws a Bode plot displaying
the variation of phase and gain with different frequencies. Click or tap the function
navigation icon
and select Bode Plots to enter the "Bode Plots" setting menu.

Figure 15.1 "Bode Plots" Setting Menu

Click or tap
to minimize the "Bode Plots" menu to simplify the display, as
shown in the figure below.

Figure 15.2 "Bode Plots" Setting Menu-Simplified Mode

www.calcert.com
15
10
20
5
25
0
30
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.