Recommended Equipment; Pmd Activation And Self Calibration; Configuring The Pmd Test - Viavi ODM V2 User Manual

Optical dispersion measurement module
Table of Contents

Advertisement

Chapter 2 Polarization Mode Dispersion Measurement

Recommended equipment

Recommended equipment

To perform a PMD measurement, the following equipment is recommended:
ONA-800 with a module as mentioned above, and associated optical connectors.
OBS-5XX Optical Broadband Source, or 81BBS2A Optical broadband source
modules.
Fiber inspection scope with associated optical connector tips.
Cleaning kit.
Two fiber patchcords with required optical connectors.
NOTE
Method used to measure the Polarization Mode Dispersion (PMD) is
described in the Reference Guide to Fiber Optic Testing - Vol2.
NOTE
The PMD value obtained by the fix analyzer method is the mean DGD value,
also designated as «PMD value».

PMD Activation and self calibration

1
Press H
2
Select the ODM and press the
The PMD results page displays and the auto-calibration starts automatically.
A bargraph
at the bottom of the screen. Wait for the calibration to be fully completed before
continuing.

Configuring the PMD test

To access the PMD test setup menu, press the S
You can choose the default values by pressing the Auto-set key or define your own
configuration.
8
> Test.
OME
«self calibration»
User Manual
PMD
function
informs of the progression status of the calibration
button on the ONA-800.
ETUP
700MAN902 Rev. 000
.

Advertisement

Table of Contents
loading

Table of Contents