The flexed and partially loaded knee will also disable the stance phase on slopes and ramps to
allow for greater knee flexion and more ground clearance in the swing phase.

9 MyModes

The O&P professional can activate and configure up to five MyModes in addition to basic mode
using adjustment software. These can be selected by using the Cockpit app. Only the first
three MyModes can be selected using movement patterns. Switching by using movement pat­
terns has to be activated in the adjustment software by the O&P professional.
These modes are intended for specific motion patterns or postures (e.g. inline skating, running
(jogging) ...). Settings can be adjusted using the Cockpit app (see page 80).

9.1 Running function as configured MyMode

INFORMATION
The running function will work with specialised running feet such as the 1E95 Challenger as
well as with prosthetic feet with axial compression such as the 1C61 Triton Vertical Shock. For
details on assembly and alignment please refer to the instructions for use of the foot.
Feet without axial compression are generally not suited for running.

9.2 Switching MyModes with the cockpit app

INFORMATION
Bluetooth on the prosthesis must be turned on in order to use the Cockpit app.
If Bluetooth is switched off, it can be turned on by turning the prosthesis upside-down (function
only available in basic mode) or by connecting/disconnecting the battery charger. Bluetooth is
then turned on for approx. 2 minutes. During this time, the app must be started and used to
establish a connection. If required, Bluetooth on the prosthesis can be switched on permanently
afterwards (see page 81).
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app (see page 78) or if mute mode (silent
mode) is activated, there are no beep signals.
Once a connection to a prosthesis has been established, the cockpit app can be used to switch
between the MyModes.
84
For running over longer periods of time, the O&P professional can configure a
"Running" MyMode which can be activated using the cockpit app or a move­
ment pattern.
In this mode, every step will be performed as a running step with larger swing
phase angle and no preflexion at heel strike (PreFlex) (see page 83).