4.5 Battery charger information

WARNING
Storing/transporting the product near active implanted systems
Interference with active implantable systems (e.g. pacemaker, defibrillator, etc.) due to the
product's magnetic field.
When storing/transporting the product in the immediate vicinity of active implantable sys­
tems, ensure that the minimum distances stipulated by the manufacturer of the implant are
observed.
Make sure to observe any operating conditions and safety notices stipulated by the manufac­
turer of the implant.
NOTICE
Penetration of dirt and humidity into the product
Lack of proper charging functionality due to malfunction.
Ensure that neither solid particles nor liquids can penetrate into the product.
NOTICE
Mechanical stress on the power supply/battery charger
Lack of proper charging functionality due to malfunction.
Do not subject the power supply/battery charger to mechanical vibrations or impacts.
Check the power supply/battery charger for visible damage before each use.
NOTICE
Operating the power supply unit/charger outside of the permissible temperature range
Lack of proper charging functionality due to malfunction.
Only use the power supply unit/charger for charging within the allowable temperature range.
The section "Technical data" contains information on the allowable temperature range (see
page 90).
NOTICE
Independent changes or modifications carried out to the battery charger
Lack of proper charging functionality due to malfunction.
Have any changes or modifications carried out only by Ottobock authorised, qualified per­
sonnel.
NOTICE
Contact of the battery charger with magnetic data storage devices
Wiping of the data storage device.
Do not place the battery charger on credit cards, diskettes, audio or video cassettes.
NOTICE
Improper care of the housing
Damage to the casing through the use of acetone, white spirit or similar solvents.
Only clean the housing with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der­
maClean).