Specifications

Channel skew

Skew adjust
Pattern shift
Nominal
channel-to-channel pattern
skew

Data error insertion

Error insertion types
Error insertion rate

Control interfaces

Front panel touchscreen
Computer programmable
interface
22
The channel skew is the timing of the data outputs.
Relative to nominal position
PPG1600 and PPG3000. Range = ±50 ps
Resolution = 100 fs
PPG3200. Range = ±25 ps
Resolution = 100 fs
Advance or delay. This is equivalent to unlimited shifting since this range allows
shifting the longest pattern to any position.
Range. ±(2
-1)
30
Resolution. 1 bit
< ±2 UI, Time difference between patterns on a 2-channel PPG3000 series, skew
adjust and bit shift at 0.
Single or rate-based
Range. 1 x 10
-3
Resolution. 3 digits
Yes, edit all instrument settings.
GUI
USB TMC, program all instrument settings.
to 1 x 10
BER
-15
PPG1600, PPG3000, & PPG3200 Pattern Generator User Manual