Specifications
Termination voltage range
Data output jitter
Connector type
Output impedance
PPG3000 & PPG1600 data outputs
Amplitude range
Offset range
14
-2.0 V to +3.3 V window. Programmable/adjustable. Applied by user via 50 Ω.
This setting is used in cases where the load being driven is terminated at a level
other than zero volts. The effect of the termination voltage on the output voltage is
shown in the following figure. To ensure proper operation, never load the output
with a termination voltage less than Voh minus 3 V.
250 fs
RJ typical at 32 Gb/s using PRBS 2
RMS
2.4 mm
50 Ω. Single-ended
100 Ω. Differential
250 mV to 2.0 V. Single-ended
500 mV to 4.0 V. Differential. Each positive and negative differential output is
independently programmable.
-2 V to +3.0 V window. Programmable/adjustable.
PPG1600, PPG3000, & PPG3200 Pattern Generator User Manual
-1 pattern
11