When rack mounting the instrument, ensure there is adequate airflow around the
instrument rear, sides, and bottom to ensure proper cooling. Adequate airflow
enables air temperatures within approximately one inch of the instrument surfaces
to remain within specified limits under all operating conditions.
The following table describes the maximum operating environmental ratings for
your instrument.
Table 2: Maximum operating environmental considerations
Feature
Description
Temperature
40 °C (104 °F)
Humidity
80% for temperatures up to 31 °C (88 °F) decreasing linearly to 50%
at 40 °C (104 °F)
Altitude
2000 m (6562 ft.)