Nimbus/DFS Service Manual
5.5
Fold back the cell assembly at the foot end of the bed to reveal the sensor pad hoses
(Fig 41, Item 1). Connect the sensor pad hoses to the hoses on the back of the CPR unit.
Note: The sensor pad hoses for Nimbus 2/DFS 2 assemblies have to be routed through the hole in the
base of the loop sheet and connected to the CPR Unit through the manifold assembly.
5.6
Insert the end of the CPR unit (Fig 42, Item 1) into the pockets (Fig 42, Item 2) on the inside of
the base cover. Push the CPR unit into the hole in the base cover.
5.7
Pass the CPR retaining strap (Fig 42, Item 3) around the back of the CPR unit and up between
the manifold of the cell assembly and the CPR unit.
5.8
Connect the pop studs (Fig 34, Item 4) on the loop to secure the CPR unit.
5.9
Pass the end of the heel protector (Fig 34, Item 2) thought the loop (Fig 34, Item 3) and connect
the pop stud (Fig 34, Item 1) to the inside of the base cover.
5.10
Connect the three (six on Nimbus 2/DFS 2) pop studs, which attach the cell loop sheet to the
base cover, at the foot end of the bed (Refer to Figure 36).
5.11
Connect the six pop studs, which attach the cell loop sheet to the base cover, at the head end of
the bed (Refer to Figure 35 ).
5.12
Install the top cover (Refer to Page 3, Section 3).
6
Replacing the Soft Foam Sheet
6.1
Remove the top cover (Refer to Page 3, Section 2).
6.2
Disconnect the three (six on Nimbus 2/DFS 2) pop studs (Fig 35, Item 1) which connect the cell
loop sheet to the base cover at the foot end of the bed.
6.3
Disconnect the six pop studs (Fig 36, Item 1) which connect the cell loop sheet to the base
cover at the head end of the bed.
6.4
Lift the mattress cell assembly from the side opposite to the manifold and remove the soft foam
sheet (Fig 37, Item 1) from the corner pockets (Fig 37, Item 3) in the base cover and pull the
foam sheet from around the sensor pad hoses using the slit in the end of the foam sheet.
6.5
Carefully remove the soft foam sheet so as to not damage the sensor pad underneath.
6.6
Lift the mattress cell assembly from the side opposite to the manifold and position the soft foam
sheet (Fig 37, Item 1) over the sensor pad. Fit the foam sheet around the sensor pad hoses
using the slit in the end of the foam sheet. Insert the corners of the foam sheet into the corner
pockets (Fig 37, Item 3) in the base cover.
6.7
Connect the three (six on Nimbus 2/DFS 2) pop studs, which attach the cell loop sheet to the
base cover, at the foot end of the bed (Refer to Figure 36).
6.8
Connect the six pop studs, which attach the cell loop sheet to the base cover, at the head end of
the bed (Refer to Figure 35).
6.9
Install the top cover (Refer to Page 3, Section 3).
Issue 2
April 2010
SER0001
Mattress Repair
Chap 4
Page 9
Need help?
Do you have a question about the Arjohuntleigh NIMBUS II and is the answer not in the manual?
Questions and answers