GENERAL SAFETY Before you connect the system pump to a mains socket, read carefully all the installation instructions contained within this manual. The system has been designed to comply with regulatory safety standards including: • EN60601-1:1990/A13:1996 and IEC 60601-1:1988/A2:1995 • UL60601-1, UL2601-1 and CAN/CSA C22.2 No. 601.1-M90 Safety Warnings •...
Page 6
• Placing extra layers between the patient and the mattress potentially reduces the benefits provided by the mattress and should be avoided or kept to a minimum. As part of sensible pressure area care, it is advisable to avoid wearing clothing which may cause areas of localised high pressure due to creases, seams, etc.
1. Introduction ® About this Manual This manual is your introduction to the Nimbus systems. Use it to initially set up Nimbus 3 Professional the system, and keep it as a reference for day-to-day routines and as a guide to maintenance. About Nimbus 3 and Nimbus 3 Professional are Dynamic Nimbus 3 and Nimbus 3 Professional...
Nimbus 3 Pump The same pump is used on the Nimbus 3 Nimbus 3 systems. Professional The pump comprises a moulded case with non-slip feet on the base and rear, and an integral carry handle. Mains Power Switch Carry Handle Alarm Indicators mm Hg mmHg...
Nimbus 3 Mattress The mattress comprises the following Nimbus 3 components: Detachable Cover 3 Head Cells 4 Thigh Cells Head End 5 “Heelguard” Cells Drag Handle 8 Torso Cells Securing Strap Foot End Carry Handle CPR Control Pump Tubeset Transport Control The standard protective cover comprises a 2-way stretch Detachable Cover cover zipped to a durable anti-slip base.
The advanced sensor pad is under the cells, AutoMatt AutoMatt and makes sure that the patient is automatically supported at optimum pressures regardless of size, height, position or weight distribution. The CPR (Cardio-Pulmonary Resuscitation) Control is CPR Control at the foot end of the mattress, and allows the air to be evacuated in under 10 seconds.
Page 11
This is a two-position rotary-action control at the head Head Section Deflate Control end of the mattress: • . The three cells in the Head Dynamic (Normal) Mode Section are inflated at a constant pressure and the remaining 17 cells alternate. •...
2. Clinical Applications Indications The systems are Nimbus 3 Nimbus 3 Professional indicated for the prevention and management of all categories of pressure ulcer when combined with an individualised monitoring, repositioning and wound care programme. mattress is Nimbus 3 Nimbus 3 Professional designed for patients weighing up to 250 kg (550 lb) .
3. Installation The Nimbus 3 and Nimbus 3 Professional systems are very simple to install using the following guidelines. Refer to Section 4, Page 11 “Controls, Alarms and Indicators” for a comprehensive description of the controls and indicators on the pump and mattress.
Page 14
4. For mattresses only, check the Nimbus 3 AutoMatt sensor pad, as follows: • Unzip the cover on one side of the mattress only. • Pull the side of the mattress away from the cells. • The sensor pad is situated under the cells AutoMatt between the soft and hard foam sheets.
1. Make sure that all 16 Vent Valves are closed. Additional Checks on the Nimbus 3 Professional Mattress Closed Open 2. Make sure that the Head Section Deflate Control is set to Dynamic (Normal) Mode TriCell Head Section Deflate Dynamic (Normal) Mode Installing the Pump 1.
Connecting the To connect the tubeset to the mattress and pump: Tubeset 1. Locate the bottom of the tubeset connector onto the bottom of the pump/mattress connector. 2. Pull the top of the tubeset connector up and over the top of the pump/mattress connector, until the tubeset connector “clicks”...
Pump Indicators The pump front panel has the following indicators: The green indicator below the ON / RESET ALARM ON / RESET ALARM switch is illuminated when the mains power is POWER connected and the pump switched on. The indicator on the button is illuminated when STATIC Mode STATIC...
Page 19
(Refer to “LOW PRESSURE” on • LOW PRESSURE page 12). (Refer to “HIGH PRESSURE” on • HIGH PRESSURE page 12). (Refer to “PUMP FAULT” on • PUMP FAULT page 13). • (Refer to “POWER Fail” on page 13). POWER For all alarm conditions except , once the alarm condition has Power Fail been detected and displayed, it can only be cancelled by switching the pump...
Mattress Controls All mattresses Nimbus 3 Nimbus 3 Professional have the following two controls, situated at the foot end of the mattress: NORMAL FAST DEFLATE NORMAL TRANSPORT TRANSPORT CPR Control Transport Control This sets the mattress into mode where the Transport Control TRANSPORT support surface is equally pressurised and the pump and...
Page 21
pressure to support the patient’s shoulders. The remaining 16 cells alternate. TriCell Head Section Deflate Dynamic (Normal) Mode The seven Torso cells, four Thigh cells and five 16 Vent Valves cells have individual Vent Valves to allow Heelguard each cell to be independently deflated, to assist with pressure area and patient care management.
5. Operation These instructions cover day-to-day operation of the system. Other operations, such as maintenance and repair, should only be carried out by suitably qualified personnel. Refer to Section 4, Page 11 “Controls, Alarms and Indicators” for a comprehensive description of the controls and indicators on the pump and mattress.
The battery may have become discharged or reached the end of its life. It is therefore recommended that the alarm is tested before the pump is used, as follows. 1. Connect the pump to the mains power supply, switch and allow it to run for 10-15 seconds. 2.
System Optimisation The systems Nimbus 3 Nimbus 3 Professional automatically compensate for patient weight distribution and position, to optimise the pressure relieving performance. To make sure that the pressure relieving properties are not impaired, the mattress cover must not be pulled tight and covering sheets should fit loosely using the attached clips.
To change the operating mode: 1. To select mode from mode press the Static Dynamic button once. An audible tone will sound STATIC and the indicator on the button will illuminate to show that the system is in mode. Static 2.
CPR Control IMPORTANT IN THE EVENT OF CARDIAC ARREST. In the event of a patient suffering cardiac arrest and CPR needing to be administered: 1. Lift the red CPR handle at the foot end of the To Activate the CPR mattress.
Patient Positioning Guidance for the Nimbus 3 Professional Mattress mattress allows the patient to be placed in either the Nimbus 3 Professional Supine or Prone positions. WARNING A full patient assessment, as to the suitability for Prone Nursing, is essential before commencing the procedure. Safety sides should be used where appropriate (Refer to “General Safety”...
Page 28
3. Prone Position. Head Section Cells Cell 4 (1 - 3) Fully Deflated 4. It is recommended that a minimum of four staff will be required to turn the patient from the Supine to the Prone position. • The anaesthetist or most senior member of the team should be positioned at the head end of the bed and will co-ordinate the turning procedure.
6. Decontamination The following processes are recommended, but should be adapted to comply with the local or national guidelines (Decontamination of Medical Devices) which may apply within the Healthcare Facility or the country of use. If you are uncertain, you should seek advice from your local Infection Control Specialist.
Page 30
DO NOT WRING/MANGLE, AUTOCLAVE OR USE PHENOLIC BASED SOLUTIONS. For information for the mattress top cover, including Thermal Disinfection laundering guidelines, refer to “Cover Specification” on page 30.
7. Routine Maintenance Nimbus 3 and Nimbus 3 Professional Systems The equipment has been designed to be virtually Maintenance maintenance-free between service periods. ArjoHuntleigh will make available on request service Servicing manuals, component parts lists and other information necessary for ArjoHuntleigh trained personnel to repair the system.
Serial Number Labels The serial number label for the pump is on the back of Pump the pump case. The serial number label for the mattress is on the top of Mattress the CPR/Transport Control, on the outside of the mattress at the foot end.
8. Troubleshooting The following table provides a troubleshooting guide for the Nimbus 3 systems in the event of malfunction. Nimbus 3 Professional Refer to Section 4, Page 11 “Controls, Alarms and Indicators” for a comprehensive description of the alarms and indicators on the pump. Indicator Possible Cause Remedy...
9. Technical Description PUMP Model: Nimbus 3 / Nimbus 3 Professional Part Numbers: 151033 Supply Voltage: 120 VAC Supply Frequency: 60 Hz Power Input: 35 VA Size: 20 x 8.7 x 4” (508 x 220 x 100 mm) Weight: 12.5 lb (5.7 kg) Case Material: ABS Plastic...
COVER SPECIFICATION Standard Cover ® ® Feature Advantex eVENT Fabric ® (Dartex) Removable Cover Moisture Vapour 12 times higher Permeable Air Permeable Low Friction 18% lower 20% lower Water Resistant / Repellent Infection Material coating is Material coating is INERT MATERIAL Control Bacteriostatic, Bacteriostatic,...
CLEANING SYMBOLS Wash at 160°F (71°C) for a minimum Tumble dry at 176-185°F 80-85°C) of 3 minutes Wash at 95°C (203°F) for a minimum Tumble dry at 130°C of 15 minutes Wash at 149°F (65°C) for a minimum Wipe surface with damp cloth of 10 minutes Do not iron Do Not Use Phenol-based cleaning Solutions...
Page 39
AUSTRALIA ITALY SWITZERLAND ArjoHuntleigh Pty Ltd ArjoHuntleigh S.p.A. ArjoHuntleigh AG PO Box 330 Via Tor Vergata, 432 Florenzstrasse 1D Hamilton Hill IT-ROMA 00133 CH-BASEL 4023 AU-6963 WESTERN T: +39 06-87426214 T: +41 (0) 61 337 97 77 AUSTRALIA F: +39 06-87426222 F: +41 (0) 61 311 97 42 T: +61 8 9 337 4111 F: +61 8 9 337 9077...