Removing the Bio Filter (Nimbus 3 and DFS 3 Pumps Only)
47.1
Disassemble the Rear Case (Refer to Chapter 3, Page 5, Section 4).
47.2
If it is necessary to replace the Bio Filter hoses as well as the Bio Filter, then cut and remove the
cable ties as shown in Figure 32.
47.3
Ease the hoses off and remove the Bio Filter (Fig 32, Item 10). If necessary remove the hoses
and the corner adaptor.
48
Installing the Bio Filter (Nimbus 3 and DFS 3 Pumps Only)
48.1
If the hoses are being replaced then cut new lengths of hose using the old hoses as a template.
48.2
Push the new Bio Filter (Fig 32, Item 10) onto the hose which leads to the Silencer Bag. Ensure
the flow indicator on the Bio Filter is uppermost and points towards the Silencer Bag hose.
48.3
Install the remaining hoses and corner adaptor to the Bio Filter and the Compressor.
48.4
Cable tie the hose connection to the Compressor.
48.5
Cable tie the Bio Filter hose to the Power Switch to Fuse Holder Wires.
48.6
Assemble the rear case (Refer to Page 5, Section 5).
48.7
Fit a new Bio Filter Label to the rear case of the pump.
Issue 2
April 2010
SER0001
Pump Repair
Chap 3
Page 59