Replacing The Mains/Power Switch - Getinge Arjohuntleigh NIMBUS II Service Manual

Table of Contents

Advertisement

Nimbus/DFS Service Manual
43

Replacing the Mains/Power Switch

WARNING: BEFORE REPLACING THE MAINS/POWER SWITCH, ENSURE THE PUMP
UNIT HAS BEEN ISOLATED FROM THE MAINS/POWER SUPPLY BY
REMOVING THE MAINS/POWER PLUG FROM THE WALL SOCKET.
43.1
Disassemble the Rear Case (Refer to Chapter 3, Page 5, Section 4).
43.2
Disconnect the connectors from the back of the Power Switch (Refer to Figure 11, 12 13 or 14 -
depending on the PCB fitted to the pump).
43.3
Compress the clips on the top and bottom of the switch on the inside of the pump casing and
push the switch out of the front panel.
43.4
Install the new Power Switch from the front of the pump with the circle uppermost, as shown in
Figure 28. Push the Switch until it clips into the front panel.
43.5
Connect the connectors to the back of the Power Switch (Refer to Figure 11, 12 13 or 14 -
depending on the PCB fitted to the pump).
43.6
Assemble the rear case (Refer to Page 5, Section 5).
Issue 2
April 2010
SER0001
Pump Repair
Chap 3
Page 53

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Arjohuntleigh NIMBUS II and is the answer not in the manual?

This manual is also suitable for:

Arjohuntleigh nimbus 3Arjohuntleigh nimbus 3 professionalArjohuntleigh dfs2Arjohuntleigh dfs3

Table of Contents