Mattress Repair
19
Removing the CPR Unit
19.1
Remove the top cover (Refer to Page 3, Section 2).
19.2
Disconnect the three pop studs (Fig 35, Item 1) which connect the cell loop sheet to the base
cover at the foot end of the bed.
19.3
Unclip the pop stud (Fig 34, Item 1) from the heel protector (Fig 34, Item 2) at the foot end of the
bed.
19.4
Pull the heel protector (Fig 34, Item 2) out of the loop (Fig 34, Item 3) and unclip the two pop
studs (Fig 34, Item 4).
19.5
Push the transport knob end of the CPR unit out of the base cover and then remove the CPR
unit from the two corner pockets.
19.6
Fold back the cell assembly at the foot end of the bed to reveal the sensor pad hoses
(Fig 41, Item 1). Disconnect the sensor pad hoses from the hoses on the back of the CPR unit.
19.7
Disconnect the heel protector tube (Fig 46, Item 1) from the manifold.
19.8
Remove the clip (Fig 46, Item 2) and disconnect the blank connector from the manifold.
19.9
Disconnect the three CPR connectors (Fig 46, Item 3) from the manifold assembly.
19.10 Remove the CPR Unit.
20
Installing the CPR Unit
20.1
Connect the three CPR connectors (Fig 46, Item 3) to the manifold assembly.
20.2
Connect the blank connector through the hole in the CPR unit and the clip and into the manifold
assembly. Tighten the clip.
20.3
Connect the heel protector tube (Fig 46, Item 1) to the manifold.
20.4
Fold back the cell assembly at the foot end of the bed to reveal the sensor pad hoses
(Fig 41, Item 1). Connect the sensor pad hoses from the hoses on the back of the CPR unit.
20.5
Connect the three pop studs, which attach the cell loop sheet to the base cover, at the head end
of the bed (Refer to Figure 35).
20.6
Insert the end of the CPR unit (Fig 42, Item 1) into the pockets (Fig 42, Item 2) on the inside of
the base cover. Push the CPR unit into the hole in the base cover.
20.7
Pass the CPR retaining strap (Fig 42, Item 3) around the back of the CPR unit and up between
the manifold of the cell assembly and the CPR unit.
20.8
Connect the pop studs (Fig 34, Item 4) on the loop to secure the CPR unit.
20.9
Pass the end of the heel protector (Fig 34, Item 2) thought the loop (Fig 34, Item 3) and connect
the pop stud (Fig , Item 1) to the inside of the base cover.
20.10 Install the top cover (Refer to Page 3, Section 3).
Chap 4
Page 18
SER0001
Nimbus/DFS Service Manual
Issue 2
April 2010
Need help?