Advanced settings

View signal strength bar?

12-8
NELLCOR® OXIMAX® SpO2: SpO2 Menu
NOTE: The auto-set option is not available for SpO
set to set the alarm limits for oxygen saturation. However, when SpO
the source for HR/Pulse, using the auto-set option in the Setup HR/
Pulse menu will set the alarm limits around the SpO
SpO
Adult, pediatric and
neonatal
Adjustable
Heart
Rate/Pulse
Low
adult,
30 - 245 (bpm)
pediatric,
neonatal
*Only when SpO
is the source for HR/Pulse or the View SpO2 PR? option is set to Yes.
2
This option lets you view the signal strength bar, which is an indication of the
strength of the electronic signal produced by the photodiode. A stronger signal
makes more bars visible, and a weaker signal makes fewer bars visible. The size
of the signal strength bar is proportional to strength of the peripheral pulsations.
It does not show changes in percentage of oxygen saturation.
Dash 2500 Patient Monitor
Alarm Limits Saturation: %
2
Low
20% - 95%
SpO
Alarm Limits: PR*
2
choice: 20%, 30%, 40%
High
Low
35 - 250 (bpm) current reading
minus chosen
percentage
You cannot choose auto-
2.
2
PR value.
2
High
25% - 100%
Auto-set
High
current reading
plus chosen
percentage