Alarms

Name of Alarm
SpO
High
2
SpO
Low
2
SpO
PR High*
2
SpO
PR Low*
2
SpO
Sensor
2
Disconnected
SpO
Lost Pulse
2
SpO
Sensor Faulty
2
SpO
Sensor Off
2
* Not active when the View SpO2 PR? option is set to no.
2042480-001 C
NELLCOR® OXIMAX® SpO2: Alarms
SpO
Alarms
2
Cause
value is greater than "hi"
alarm limit
value is less than "lo" alarm
limit
value is greater than "hi"
alarm limit as set in HR/
Pulse
value is less than "lo" alarm
limit as set in HR/Pulse
sensor disconnected from
module
no or very low SpO
signal,
2
signal was lost
SpO
sensor or cable
2
possibly defective, cable
not connected properly
SpO
sensor has been
2
removed from the patient
Dash 2500 Patient Monitor
Response
message appears in alarm message
field; parameter label, violated limit (if
limits set to display), and value in
violation flashes in vital signs field
message appears in alarm message
field; parameter label, PR label,
violated limit, violated PR limit (if limits
set to display), and value in violation
flashes in vital signs field
message appears in waveform and
alarm message field; parameter label
flashes in vital signs field
message appears in waveform field,
alarm message field, and in SelectBox
when Enable spot check is set to yes;
parameter label flashes in vital signs
field
message appears in waveform and
alarm message field; parameter label
flashes in vital signs field
message appears in waveform field,
alarm message field, and in SelectBox
when Enable spot check is set to yes;
parameter label flashes in vital signs
field
12-17