Page 1
Dash Responder Version V 1.0 Servicing Instructions 2004768-001 ENG Revision F...
Page 2
During repairs/service interventions, observe the protective measures against damage due to ESD. • GE Medical Systems Information Technologies GmbH is responsible for the effects on safety, reliability, and performance of the device, only if − assembly operations, extensions, readjustments, modifications, or...
Configuration of Instrument Part Numbers __________________________________________ 9 Configuration of the PCB Part Numbers ____________________________________________ 9 Instrument status documentation __________________________________________________ 9 Instrument Versions of the Dash Responder __________________________________ 9 General Description_______________________________________________10 Dash Responder Block Diagram ___________________________________________ 10 Dash Responder Device___________________________________________________ 10...
Page 4
Operating Position_______________________________________________________ 52 Type of Protection _______________________________________________________ 52 Environment ___________________________________________________________ 52 Operation ___________________________________________________________________ 52 Transport and Storage _________________________________________________________ 53 Dimensions _____________________________________________________________ 53 Weight ________________________________________________________________ 53 Environmental Tests _____________________________________________________ 53 Servicing Manual Dash Responder V1.0 Revision F Page 4 of 72 2004768-001...
Page 5
Battery conditioning test _______________________________________________________ 59 Electrical Safety Tests ___________________________________________61 Patient (Sink) Leakage Current Test________________________________________ 62 Spare Parts List ________________________________________________64 Master Record Index (see Appendix) _______________________________69 Circuit Diagrams (see Appendix)___________________________________71 Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 5 of 72...
Page 6
Languages of Dash Responder Version 1 Revision History This manual is subject to the GE Medical Systems Information Technologies change order service. The revision code, a letter that follows the document part number, changes with every update of the manual.
When the devices are on, the battery will charge only when the Dash Responder does not require the full energy for operation the patient monitor is able to deliver.
Page 8
Contacts for test discharge (possible only with adapter lead for adhesive defibrillation pads) Connector for exchange of the defibrillation electrodes (switch off the device before exchanging the electrodes!) Servicing Manual Dash Responder V1.0 Revision F Page 8 of 72 2004768-001...
Master Record Index 2002550-013 The master record index lists the valid configurations for the Dash Responder. Numbers for all permissible components of the device are listed. Spare part numbers are provided where available. Instrument Versions of the Dash Responder...
• Shock delivering and measuring of applied energy. The PIC-controllers initiate the switching of the relais to deliver the energy to the patient according Servicing Manual Dash Responder V1.0 Revision F Page 10 of 72 2004768-001...
Any pressed keys are decoded by the PCB Digital and the LEDs on the keypad get illuminated under control of the PCB Digital. • Sound generation The alarm sounds are generated and amplified on the PCB Digital by a sound chip. Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 11 of 72...
The battery gets charged if it needs to be and if energy is provided by a monitor connected to the Dash Responder. The PCB Powermanagement powers the device and guarantees that every remaining energy from the monitor is used to charge the battery, but the drawn current does not exceed the maximum current the monitor is able to deliver.
The generated voltage level is sufficient to deliver any selected current level to the highest expectable patient resistance. • Switched current source for pace pulses The pace pulse current is regulated by means of a programmable current source. Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 13 of 72...
PCB Digital Battery The battery in the Dash Responder is a rechargeable NiCd battery with a charge level of 2000 mAh and nominal 12 Volt (10 cells). The battery contains a charge level monitoring system to measure the remaining charge in the battery.
Start of Conditioning • Connect the monitor-defibrillator system to the power line. • Turn the defibrillator on. Servicing Manual Dash Responder V1.0 Revision F Page 16 of 72 2004768-001...
HARGE key. These are the steps of the conditioning program: the Dash Responder switches off, the battery charges (up to about seven hours, depending on the initial energy level), the green LED blinks, the yellow LED is solid on the battery discharges (seven hours), the green LED blinks, the yellow LED is off...
Dash Responder. Software Download Information The product software for the Dash Responder can be loaded into the flash memory by means of a PC or laptop. This requires the Dash Responder Service kit, a complete set for download, calibration and test.
OWNLOAD After successful installation of the software (duration about 3 minutes), the message ‘E ’ appears in the bottom line of the service tool window. The Dash Responder restarts. DOWNLOAD Enter the previous setup (see User configurable settings on page 23).
The charge level of the battery should be sufficient for several shocks. The following steps describe how to calibrate the Dash Responder Servicing Manual Dash Responder V1.0 Revision F...
GE Medical Systems Information Technologies Connect the DRST-Box to the Dash Responder and the PC (COM 1 - 4). Check at the DRST-Box that • the LED 'P ' is ON and RESENT • the Switch '9-18V' is OFF for software version V 1.0 •...
GE Medical Systems Information Technologies Attach the DRST-Box to the Dash Responder instead of a Dash Monitor. Connect the power supply and the serial cable to the DRST-Box. The other end of the serial cable should fit into a free COM-port of your PC. You may need a standard 9pin/25pin adapter if the PC has a 25-pin socket COM.
NERGY ELECT two seconds. User configurable settings Some settings of the Dash Responder are configurable to meet the user’s requirements. The settings can be made in the Service Menu and comprise following values: • Volume of the alarm sound <L >...
Page 24
GE Medical Systems Information Technologies Defib Autosequence Setup EVEL EVEL EVEL <200J> <200J> <360J> Pacer Setup <60 > Service Mode ERVICE OOLS Display Setup CKLIT ONTR ∇ ∆ ∇ ∆ Servicing Manual Dash Responder V1.0 Revision F Page 24 of 72 2004768-001...
Page 25
GE Medical Systems Information Technologies Language Setup ∆ NGLISH EUTSCH ∇ RANÇAIS EDERLANDS TALIANO SPAÑOL VENSKA Battery Conditioning <2346mAh> <2240mAh> <122%> <1> TART Service Tools ALIBR OWNLD Component Test Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 25 of 72...
Disconnect the two connectors to the paddle connector on the rear of the unit. One connector (PC/) is located at the left side of the air coil (L506) the other one (SHOCK_P/, SHOCK_M/) is Servicing Manual Dash Responder V1.0 Revision F...
Page 27
• Fasten the capacitor on the bracket with two cable ties (360mm * 7,5mm). Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 27 of 72...
Fix the PCB with its three screws. • Connect the flex flat cable to the keypad. This can best be done with a goose-necked tweezers. Make sure to fasten the actuator properly. Servicing Manual Dash Responder V1.0 Revision F Page 28 of 72 2004768-001...
Connect the flat ribbon cable from Digital PCB to Pacer PCB. • Make sure the flex flat cable of the monitor connector receptacle didn’t get loose during mounting the Pacer PCB. Reconnect and fasten it if necessary. Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 29 of 72...
Before reassembling the Power Management PCB the Digital PCB should be reassembled. • To reassemble the Power Management PCB the monitor connector receptacle has to be reassembled first. Servicing Manual Dash Responder V1.0 Revision F Page 30 of 72 2004768-001...
Page 31
Carefully push the battery case into its place in the upper case shell. Make sure the rubber seal around the opening of the battery case gets located in the frame properly. Be careful that the flex flat cable doesn’t get cracked. Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 31 of 72...
Thread new cable ties in the sockets for the capacitor. • Put the capacitor on the sockets in the Dash Responder and fix it with the cable ties. CAUTION Check that the Dash Responder is loaded with minimum Software Version 1.01 Servicing Manual Dash Responder V1.0...
PCB. WARNING Surface mounted components were not designed to be removed or replaced using standard soldering equipment. Removal of surface mounted components using a conventional soldering iron can Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 33 of 72...
The Service Tips listed below combined with formal training should provide the service technician with skills necessary to service and repair a defibrillator, in the event of a malfunction. Servicing Manual Dash Responder V1.0 Revision F Page 34 of 72...
Page 35
Monitor is monitor to defibrillator. power line. connected to the defibrillator Check if the monitor is properly placed on top of the defibrillator and the lever is engaged. Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 35 of 72...
ACER CTIVE pacemaker is still in operation internal pacemaker problem ACER RROR no pads are applied when pacemaker is turned on EQUIRED Servicing Manual Dash Responder V1.0 Revision F Page 36 of 72 2004768-001...
Keyboard; key pressed during device start Bootloader Flash RAM Program code Flash RAM EEPROM Failure of Internal Discharge; Caution: Risk of high voltage on HV-Capacitor Failure of high voltage control (PIC problem) Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 37 of 72...
GE Medical Systems Information Technologies Monitor Connector Described in the following table are the signals that are used by the Dash Responder; additional signal on the "Dash Hardware Peripheral Expansion Interface", which are not used for the Dash Responder are shaded.
Keyboard key line CHARGE Ground KB_DISCHARGE Keyboard key line SHOCK Ground HP_CHARGE Hardpaddle key line CHARGE Ground HP_DISCHARGE Hardpaddle key line SHOCK Ground EN_ SHOCKA Enable Defishock to PIC_A Servicing Manual Dash Responder V1.0 Revision F Page 40 of 72 2004768-001...
Page 41
Chip Select2 DISPL_EN Display Enable PWM_DISPCLK External Clock (2kHz) R/W_ Read/Write_ IO_DAT0 Data Bus0 IO_DAT1 Data Bus1 IO_DAT2 Data Bus2 IO_DAT3 Data Bus3 IO_DAT4 Data Bus4 IO_DAT5 Data Bus5 Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 41 of 72...
Page 42
Battery full (LED connected to GND) BATT_YE_LED Yellow LED is illuminated when Battery is being charged, it blinks yellow when battery charging function is defect or disturbed (LED is connected to GND) Servicing Manual Dash Responder V1.0 Revision F Page 42 of 72 2004768-001...
Page 43
GNDP Ground for VCCP power supply Ground for VCC PHASE_A_ clock for HV-Generator Ground for VCC PHASE_B_ clock for HV-Generator Ground for VCC Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 43 of 72...
Page 44
LED blinks the battery charging is erroneous. Ground KB_POWER_ON_ Keypad Power On 18V- CMOS Ground POWER_ON_KEY Level shifted Power On signal Ground POWER_OFF_ Supply switch controlled by the processor Ground Servicing Manual Dash Responder V1.0 Revision F Page 44 of 72 2004768-001...
Page 45
GNDD Ground for VCCD VCCD unregulated Power Supply for +8V...+18 PCB Digital GNDP Ground for VCCP VCCP unregulated Power Supply for +8V...+18 PCB PACER GNDP Ground for VCCP Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 45 of 72...
Page 46
Data, Command, Status Port SMCDAT_7 Data, Command, Status Port Low Voltage Detect Ground SMC_RB_ Ready/Busy_ SMC_RE_ Read Enable SMC_CE_ Chip Enable +3,3V power supply +3,3V Ground SMC_ABSENT Card Detect Servicing Manual Dash Responder V1.0 Revision F Page 46 of 72 2004768-001...
Active (measured versus GND) PACE_P High Voltage + (floating) (single) +200V during pace (+2500V during shock) PACE_M High Voltage - (floating) (single) -200V during pace (-2500V during shock) Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 47 of 72...
Page 48
Description Level Active HP_Charge Charge key from Paddle HP_Discharge Discharge key from Paddle Paddle4 Paddle code 4 Paddle6 Paddle code 6 Ground MISC Paddle code misc n.c. n.c. Servicing Manual Dash Responder V1.0 Revision F Page 48 of 72 2004768-001...
Pacer PCB Analog PCB Connector name P _P and P (see description Pacer PCB, page 47) Digital PCB Connector name P ACER (see description Pacer PCB, page 47) Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 49 of 72...
50-ohm load resistance (max. twice within 15 minutes) • 3-digit display of the delivered energy • warning on LCD when discharge circuit is interrupted (e.g. pacer electrode not applied) Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 51 of 72...
Type of Protection protected against penetration of spray water and small objects: IP 23 (battery inserted, cable connected) Environment Operation • operation under the following conditions considered to be normal: Servicing Manual Dash Responder V1.0 Revision F Page 52 of 72 2004768-001...
Safety labels and inscriptions on the device are clearly legible. General Cleaning Clean external surfaces The exterior surfaces may be cleaned with a lint-free cloth dampened with one of these approved solutions: Servicing Manual Dash Responder V1.0 Revision F Page 54 of 72 2004768-001...
Following Testing Equipment is required: • Calibrated Defitester (e.g. DNI Nevada or equivalent) • Safety Tester for measurements according to IEC601 • Pacer Testadapter (220 101 01) Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 55 of 72...
Pacer Mode ACER Pacer Pause ACER AUSE : For the test of the pacer keys press the keys at least 2 seconds. Switch the Dash Responder off and on. Servicing Manual Dash Responder V1.0 Revision F Page 56 of 72 2004768-001...
A checkered pattern will appear. • Verify that no pixels, rows or columns are missing. Detection of internal Electrodes • Connect the internal Defi Electrode (217 308 01) to the Dash Responder. Increase energy with key. NERGY ELECT • Check if the energy selection is limited to 50J.
Using the +P/min; -P/min key select any frequency. • Using the oscilloscope measure the drop in voltage across the internal measuring resistor. Adjust the following values on the Dash Responder and verify the values measured by the oscilloscope: DASH-Responder setup Values to measure mA ±...
20% or ± 20 mA) Defib Function Test and Pacer Option Test The following tests are performed with the Service Tool (2006861-001). If the Dash Responder is connected with a Dash Monitor this tests could also be done by means of the Dash Monitor.
Page 60
While the battery is being charged, the yellow C LED is solid illuminated, it is off HARGING TATUS during discharging the battery and it is blinking if there is a charging problem. Servicing Manual Dash Responder V1.0 Revision F Page 60 of 72 2004768-001...
Electrical safety tests provide a method of determining if potential electrical health hazards to the patient or operator of the device exist. To help you establish a systematic maintenance routine, GE Medical Systems recommends that you perform all safety tests presented in this chapter •...
50 µA (0.05 volts on the DMM) at 120 – 240 VAC with internal paddle electrodes. The 50 µA limit is according to ANSI/AAMI DF2. The leakage current is measured from each paddle. Servicing Manual Dash Responder V1.0 Revision F...
Page 63
GE Medical Systems Information Technologies Set the power switch on the leakage tester to OFF. Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 63 of 72...
Plastic Yoke for directing Locking Handle 2003414-001 Slide Plastic Slide for front or back 2001787-001 Spring Leg Spring Leg (metal spring) for plastic slide 2006843-001 Bracket Metal Bracket on top side. 2001788-001 Servicing Manual Dash Responder V1.0 Revision F Page 64 of 72 2004768-001...
Page 65
2005068-001 Label Label (Battery Type) Spanish 2005069-001 Label Label (Battery Type) Dutch 2005070-001 Label Label (Battery Type) Swedish 2005071-001 Plate Mount GCX Mounting Plate for GCX mounting 420001-001 Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 65 of 72...
Page 66
2007052-001 and Front Label have to be ordered separately Rubber Gasket Rubber Gasket for Front Case, order 70cm of length 2004898-001 Label Label for Dash Responder without Pacer English 2003287-001 Label Label for Dash Responder without Pacer German 2003288-001 Label...
Page 67
(external) Tools Part Description Detailed Part Description Part No. Kit Service Tool Service Tool Kit for Dash 2000 and Dash Responder, 2006861-001 including Power supply 2000300-001, DRST Box 2005378, Software Disks 2006862-001, Connection Cable 2233603. Kit Upgrade Dash Dash Responder Software V1. Kit is replaced by...
Page 68
2002853-004 Operator Manual V1 Operator Manual Version 1 Spanish 2002853-005 Operator Manual V1 Operator Manual Version 1 Dutch 2002853-006 Operator Manual V1 Operator Manual Version 1 Swedish 2002853-007 Servicing Manual Dash Responder V1.0 Revision F Page 68 of 72 2004768-001...
GE Medical Systems Information Technologies 14 Master Record Index (see Appendix) Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 69 of 72...
Page 70
GE Medical Systems Information Technologies Servicing Manual Dash Responder V1.0 Revision F Page 70 of 72 2004768-001...
GE Medical Systems Information Technologies 15 Circuit Diagrams (see Appendix) Revision F Servicing Manual Dash Responder V1.0 2004768-001 Page 71 of 72...
Page 72
GE Medical Systems Information Technologies Servicing Manual Dash Responder V1.0 Revision F Page 72 of 72 2004768-001...
Page 79
Sheet: Marquette Hellige GmbH 2002317-001 D-79007 Freiburg REVISIONS Not equipped in Version : 2002315-001 Revision-No Index Date / Name SCHEMATIC 2002316-001 ER21920 20.01.00/VRP Date / Name ER23599 17.07.00/VRP DRAWN 16.12.1999/VRP MR26708 30.10.00/VRP APPROVED 23.12.1999/WHL ANALOG DASH-ADD-ON-DEFI V. PLACKE ISSUED Thursday, September 16, 1999 $USER_CAE_B/daod_analog...
Page 80
Marquette Hellige GmbH Sheet: 2002317-002 D-79007 Freiburg REVISIONS Not equipped in Version : 2002315-001 Revision-No Index Date / Name SCHEMATIC 2002316-001 ER21920 20.01.00/VRP Date / Name ER23599 31.07.00/VRP DRAWN 16.12.1999/VRP MR26708 30.10.00/VRP APPROVED 23.12.1999/WHL ANALOG DASH-ADD-ON-DEFI V.PLACKE ISSUED $USER_CAE_B/daod_analog...
Page 81
Marquette Hellige GmbH Sheet: 2002317-003 D-79007 Freiburg REVISIONS Not equipped in Version : 20002315-001 Revision-No Index Date / Name SCHEMATIC x1 2002316-001 ER21920 20.01.00/VRP Date / Name ER23599 31.07.00/VRP DRAWN 21.12.1999/VRP MR26708 30.10.00/VRP APPROVED 23.12.1999/WHL ANALOG DASH-ADD-ON-DEFI ECO066497 28.01.01/VRP ECO067648 13.08.01/VRP V.
Page 82
Marquette Hellige GmbH Sheet: 2002317-004 D-79007 Freiburg REVISIONS Not equipped in Version : 2002315-001 Revision-No Index Date / Name SCHEMATIC 2002316-001 ER21920 20.01.00/VRP Date / Name ER23599 31.07.00/VRP DRAWN 23.12.1999/VRP MR26708 30.10.00/VRP APPROVED 23.12.1999/WHL ANALOG DASH-ADD-ON-DEFI V. PLACKE ISSUED $USER_CAE_B/daod_analog...
Page 83
Sheet: Marquette Hellige GmbH 2002317-005 D-79007 Freiburg REVISIONS Not equipped in Version : 2002315-001 Revision-No Index Date / Name SCHEMATIC 2002316-001 ER21920 20.01.00/VRP Date / Name ER23599 21.07.00/VRP DRAWN 21.12.1999/VRP MR26708 30.10.00/VRP 23.12.1999/WHL APPROVED DASH-ADD-ON-DEFI ANALOG ISSUED V. PLACKE $USER_CAE_B/daod_analog...
Page 86
Sheet: Marquette Hellige GmbH 2000924-001 D-79007 Freiburg REVISIONS Not equipped in Version : 2000922-001 Revision-No Index Date / Name SCHEMATIC 2000923-001 ER20398 12.12.99/MLM Date / Name MR25286 06.07.00/MLM DRAWN 01.09.99/MLM 03.09.99/CYS APPROVED DIGITAL DASH-ADD-ON-DEFI ISSUED M. Mueller Wednesday, December 30, 1998 $USER_CAE_A/dd_digital/design...
Page 87
GE Medical Systems 2000924-002 Sheet: D-79111 Freiburg Information Technologies GmbH Munzinger Str. 3 REVISIONS Not equipped in Version : 2000922-001 Revision-No Index Date / Name SCHEMATIC 2000923-001 ECO 065698 08.11.00/MLM Date / Name ECO 066217 09.01.01/MLM DRAWN 01.09.99/MLM ECO 066643 06.03.01/MLM...
Page 88
GE Medical Systems Sheet: 2000924-003 D-79111 Freiburg Information Technologies GmbH Munzinger Str. 3 REVISIONS Not equipped in Version : 2000922-001 Revision-No Index Date / Name SCHEMATIC 2000923-001 ER20398 08.12.99/MLM Date / Name MR25286 06.07.00/MLM DRAWN 01.09.99/MLM ECO 065698 03.09.99/CYS 08.11.00/MLM...
Page 89
Sheet: Marquette Hellige GmbH 2000924-004 D-79007 Freiburg REVISIONS Not equipped in Version : 2000922-001 Revision-No Index Date / Name SCHEMATIC 2000923-001 ER20398 03.12.99/MLM Date / Name MR25286 06.07.00/MLM DRAWN 01.09.99/MLM 03.09.99/CYS APPROVED DIGITAL DASH-ADD-ON-DEFI ISSUED M. Mueller [1999, 9, 6, 2] $USER_CAD_A/2000922-001/design/power_supply...
Page 90
Sheet: Marquette Hellige GmbH 2000924-005 D-79007 Freiburg REVISIONS Not equipped in Version : 2000922-001 Revision-No Index Date / Name SCHEMATIC 2000923-001 ER20398 03.12.99/MLM Date / Name MR25286 06.07.00/MLM 01.09.99/MLM DRAWN 03.09.99/CYS APPROVED DIGITAL DASH-ADD-ON-DEFI M. Mueller ISSUED [1999, 6, 30, 4] $USER_CAE_A/dd_dig_save1/design/CPLD...
Page 92
Sheet: Marquette Hellige GmbH 2002441-001 D-79007 Freiburg REVISIONS Not equipped in Version : 2002439-001 Revision-No Index Date / Name x1 2002440-001 ER22558 03.04.00/WHL Date / Name SCHEMATIC ER23632 04.08.00/WHL 29.03.00 / WHL DRAWN MR25491 07.02.01/WHL APPROVED 30.03.00 / GHH POWER MANAGEMENT DASH-ADD-ON-DEFI 066505 14.03.01/WHL W.LOEHNING...
Page 93
Sheet: Marquette Hellige GmbH 2002441-002 D-79007 Freiburg REVISIONS Not equipped in Version : 2002439-001 Revision-No Index Date / Name x1 2002440-001 ER22558 03.04.00/WHL Date / Name SCHEMATIC ER23632 04.08.00/WHL DRAWN 29.03.00 / WHL MR25491 07.02.01/WHL 30.03.00 / GHH APPROVED POWER MANAGEMENT DASH-ADD-ON-DEFI 066505 24.04.01/WHL ISSUED...
Page 96
Sheet: Marquette Hellige GmbH 2001199-001 D-79007 Freiburg REVISIONS Not equipped in Version : 2001197-001 Revision-No Index Date / Name SCHEMATIC 2001198-001 ER23591 01.08.00/MLM Date / Name MR26482 31.10.00/MLM DRAWN 22.10.99/MLM ECO065697 30.11.00/MLM 28.10.99/PHG APPROVED PACER DASH ADD-ON-DEFI ECO066704 05.04.01/MLM ECO066943 02.05.01/MLM ISSUED M.
Page 100
World Headquarters Asia Headquarters European Representative GE Medical Systems GE Medical Systems GE Medical Systems Information Technologies Asia; GE (China)Co., Ltd. Information Technologies, Inc. Information Technologies GmbH 8200 West Tower Avenue Munzinger Straße 3 24th Floor Shanghai MAXDO Centre, Milwaukee, WI 53223 USA...