View vitals every

Select Patient

Mini trends

View trends on main screen?

Display as

View vitals every
6-8
Trends: View Patient Trends Menu: Numbers
This menu option provides you with two ways of viewing trended information.
For continuously updated parameters, you can choose to have the data
displayed in 1 min, 5 min, 10 min, 15 min, 30 min, 1 h, 1.5 h, 2 h, 3 h intervals, or
as NIBP or TEMP. (NIBP and TEMP are considered episodic – and listed
independently – because they are not continuously updated.) Data of
parameters not continuously updated are displayed in the time interval when
the measurement was taken (e.g., NIBP and TEMP). These time intervals will also
include trends for other vital signs. This option is available when Display as is set
to numbers.
This menu option lets you view data from either the current or the previous
patient when the Save previous patient data? option in the Config settings
menu option is set at yes, and there is previous patient data available for the
viewing mode you have selected. When the Save previous patient data option
is set at no, you can view data from only the current patient.
The mini trends window is displayed at the bottom of the main monitoring
screen. It is limited to displaying only the most recent data and only that of the
Current patient. The mini trends window is an abbreviated version of the full
trends screen. Unlike the full trends screen, mini trends only displays HR/Pulse,
SpO
, NIBP, and impedance respiration (RESP). When the menu is open, you can
2
view four columns of data, and when it is closed you can view 10 columns of
data. Mini trends does not contain a NOW column. The trended information in
the mini trends window appears in either graphs or numbers mode, whichever
the default setting.
This menu option lets you turn mini trends on or off. The mini trends window
appears at the bottom of the main monitoring screen.
This option lets you choose the mode in which you view trended data in the mini
trends window. Two modes are available: numbers or graphs.
This menu option is available when the Display as option is set to numbers. It
provides you with two ways of viewing trended information in mini trends. For
continuously updated parameters, you can choose to have the data displayed
in 1 min, 5 min, 10 min, 15 min, 30min, 1 h, 1.5 h, 2 h, or 3 h intervals or as NIBP.