Advanced settings

Cardiac sweep speed

Cascade ECG?

calibration pulse

2042480-001 C
ECG: ECG Menu
You can use this option to set the sweep speed at 12.5 mm/s, 25.0 mm/s, or 50.0
mm/s. The chosen sweep speed appears next to the lead label in the waveform
field when it is a setting other than 25.0 mm/s.
NOTES
Changes you make to the sweep speed in this option will change the sweep
speed of all cardiopulmonary waveforms.
In the 5-electrode configuration module, if the sweep speed of the primary
lead waveform has been fixed at 25.0mm/s in the configuration mode, the
sweep speed of the primary lead waveform will remain at 25.0 mm/s
regardless of changes to the sweep speed of the other cardiopulmonary
waveforms.
If the Fixed ECG sweep speed option (under the Config setting menu) is set
to yes, the ECG Cardiac sweep speed menu option is not available.
This option is active in 5-electrode configuration. This option lets you view
double the number of seconds of data on the screen. When you choose yes, the
primary waveform extends to an additional field below the original field.
NOTE: Mini trends window disables the cascading feature.
When you push the Trim Knob to choose this option, one cycle of a ± 0.5 mV
square wave appears in the primary waveform.
Dash 2500 Patient Monitor
9-13