Adjust limits

Patient
adult/
pediatric
Systolic
neonatal
adult/
pediatric
Mean
neonatal
adult/
pediatric
Diastolic
neonatal
11-10
NIBP: NIBP Menu
The adjustable and auto-set alarm limits are shown in the table. Choose hi to
change the high alarm limit and lo to change the low alarm limit. You can set high
and low limits individually for systolic, diastolic, and mean values. The system
maintains appropriate separation between the high and low limits. Choose auto-
set to use the patient's current readings in setting the high and low limits.
NIBP Alarm Limits (mmHg)
Adjustable
Low
High
mmHg
30 - 285
35 - 290
> 25
< 25
30 - 135
35 - 140
> 25
< 25
20 - 255
25 - 260
> 25
< 25
20 - 120
25 - 125
> 25
< 25
10 - 215
15 - 220
> 25
< 25
10 - 105
15 - 110
> 25
< 25
Dash 2500 Patient Monitor
Auto-set
choices: 20%, 30%, or 40%
Low
current reading minus
chosen percentage
current reading minus 5
current reading minus
chosen percentage
current reading minus 5
current reading minus
chosen percentage
current reading minus 5
current reading minus
chosen percentage
current reading minus 5
current reading minus
chosen percentage
current reading minus 5
current reading minus
chosen percentage
current reading minus 5
High
current reading plus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
current reading plus
chosen percentage
current reading minus 5
2042480-001 C