GE Dash 2500 Operator's Manual page 254

Patient monitor
Hide thumbs Also See for Dash 2500:
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration – Electromagnetic Immunity
Immunity Test
EN 60601 Test Level
Electrostatic
± 6 kV contact
Discharge (ESD)
± 8 kV air
EN 61000-4-2
Electrical Fast
± 2 kV for power supply
Transient/Burst
lines
EN 61000-4-4
± 1 kV for input/output
lines
Surge
± 1 kV differential mode
EN 61000-4-5
± 2 kV common mode
< 5% U
for 0.5 cycles
Voltage dips,
< 40% U
short
for 5 cycles
interruptions
< 70% U
and voltage
variations on
for 25 cycles
power supply
input lines
EN 61000-4-11
< 5% U
for 5 s
Power
3 A/m
Frequency (50/
60 Hz) Magnetic
Field
EN 61000-4-8
E-4
EMC Compliance: Electromagnetic Compatibility (EMC)
The Dash 2500 Patient Monitor is intended for use in the electromagnetic
environment specified below. It is the responsibility of the customer or user to
assure that the Dash 2500 Patient Monitor is used in such an environment.
Compliance Level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV differential
mode
± 2 kV common
mode
(> 95% dip in U
)
< 5% U
t
t
U
) for 0.5 cycles
t
(> 60% dip in U
)
< 40% U
t
t
in U
(> 30% dip in U
)
< 70% U
t
t
in U
(> 95% dip in U
)
< 5% U
t
t
U
) for 5 s
t
3 A/m
NOTE:
U
is the AC mains voltage prior to application of the test level.
t
Dash 2500 Patient Monitor
Electromagnetic Environment – Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power should be that of a typical
commercial or hospital environment.
Mains power should be that of a typical
commercial or hospital environment.
(> 95% dip in
t
(> 60% dip
Mains power should be that of a typical
t
commercial or hospital environment. If the
) for 5 cycles
t
user of the equipment requires continued
(> 30% dip
operation during power mains interruptions,
t
it is recommended that the equipment be
) for 25 cycles
t
powered from an uninterruptible power
supply or a battery.
(> 95% dip in
t
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
2042480-001 C