View Spo2 Pr; Enable Spot Check; Turn C-Lock On; Cardiac Sweep Speed - GE Dash 2500 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 2500:
Table of Contents

Advertisement

View SpO2 PR?

Enable spot check?

Turn c-Lock on?

Cardiac sweep speed

Limit alarms priority

other alarm priorities

2042480-001 C
NELLCOR® OXIMAX® SpO2: SpO2 Menu
This option lets you view and trend the pulse rate currently being measured by
the SpO
parameter. This rate may be different from the value displayed for HR/
2
Pulse, because it is calculated from a different source (i.e., ECG). When viewable,
the SpO
PR value is also checked against the same limits set for HR/Pulse. If this
2
option is set to no, the SpO
checked against limits when it is the source for HR/Pulse.
This option lets you enable or disable the SpO
be able to take individual SpO
monitoring. If spot check is enabled, the SPOT CHECK message appears next to
the SpO
label in the waveform area. By enabling the spot check option, you can
2
apply the SpO
sensor to the patient, take a reading, then remove the sensor
2
from the patient, acknowledge the LOST PULSE or SENSOR OFF alarm, and the
Monitor returns to READY mode allowing for SpO
Choose no to continuously monitor SpO
when you remove the SpO
OFF alarm sounds indicating continuous monitoring has been interrupted. The
Monitor continues to alarm until either the patient signal is detected again or
SpO
is turned off-line.
2
This option lets you enable and disable NELLCOR
synchronization. With C-LOCK
reliable.
This option lets you set the display sweep speed to 12.5 mm/s, 25.0 mm/s, or
50.0 mm/s. Changes you make to the sweep speed in this menu will effect the
sweep speed of all cardiopulmonary waveforms with the possible exception of
the primary ECG waveform. The chosen sweep speed appears next to the SpO
label in the waveform area when it is a setting other than 25.0 mm/s.
This option lets you choose the priority of crisis or warning for the limit alarms.
The default limit alarms priority is warning.
This option lets you choose priorities for Lost pulse, Sensor disconnected,
Sensor faulty, and Sensor off alarms.
Dash 2500 Patient Monitor
PR value will only be displayed, trended, and
2
spot check option. Choose yes to
2
readings—spot checks—without continuous
2
readings at a later time.
2
. By disabling the spot check option,
2
sensor from the patient a LOST PULSE or SENSOR
2
®
C-LOCK
®
enabled, saturation measurements are more
®
ECG
2
12-9