Test Config; Modbus - Ametek 5100 HD User Manual

Tunable diode laser spectrometer
Hide thumbs Also See for 5100 HD:
Table of Contents

Advertisement

Test Config

The Test Config key is used to define device and communication settings. See
below for information on each section.

Modbus

Selects the communication setup. The Modbus protocol transmission mode
implemented is Remote Terminal Unit (RTU) with the analyzer operating as a
slave device.
Menu for Modbus key.
Slave Address
Enter the Slave Address.
Baud Rate
Select the baud rate at which you transfer data (9600 or 19200). The default
is 9600.
Time Out (ms)
Enter the time out in milliseconds.
Modbus interface parameters must be set up to establish communication. Use the
Modbus registers to obtain specific input/output data.
NOTE
Slave Address*
MODBUS
Time Out (ms)*
Modbus Registers can be found at the end of this chapter.
9600
Baud Rate
19200
|
User Interface
4-9