System Alarm/Warning bit map

System Alarm/Warning bit map
When System Status (parameter 129, 16-bit integer, modbus register 117) con-
When System Status (parameter 129, 16-bit integer, modbus register 117) contains none-zero value,
tains none-zero value, the bit map of System Alarms and System Warning (pa-
the bit map of System Alarms and System Warning (parameter 125, 126, 127, 128; 32-bit integer,
rameter 125, 126, 127, 128; 32-bit integer, modbus register 109, 110, 111, 112,
modbus register 109, 110, 111, 112, 113) contains system alarms as defined:
113) contains system alarms as defined:
Modbus
Bit
Display message
register
109
1
"TEC Hi Limit Alrm"
109
2
"TEC Lo Limit Alrm"
109
3
"TEC2 Hi Limit Alrm"
109
4
"TEC2 Lo Limit Alrm"
109
5
"Cell Hi Limit Alrm"
109
6
"Cell Lo Limit Alrm"
109
7
"Pres Hi Limit Alrm"
109
8
"Pres Lo Limit Alrm"
109
9
"Seal Hi Limit Alrm"
109
10
"Seal Lo Limit Alrm"
109
11
"File error"
109
12
"Internal error"
111
13
"TEC temp drift"
111
14
"TEC2 temp drift"
111
15
"Conc. Out of range"
111
16
"Conc. Out of range 2"
111
17
"Lowlight in Sample"
111
18
"Battery failed"
111
19
"Lowlight in Ref"
111
20
"Ref conc alarm"
111
21
"Conc. Band 1"
111
22
"Conc. Band 2"
111
23
"Conc. Band 3"
111
24
"Conc. Band 4"
111
25
"Conc. Band 1"
111
26
"Conc. Band 2"
111
27
"Conc. Band 3"
111
28
"Conc. Band 4"
111
29
"Data Invalid"
111
30
"Cell pres Hi"
111
31
"Cell pres Low"
111
32
"Span error"
112
1
"Zero error"
113
2
"User error"

Common Modbus Registers

Alarm
Related
type
Parameter ID
Alarm
181, 183, 188
Alarm
Alarm
374, 376, 381
Alarm
Alarm
192, 193, 194
Alarm
Alarm
244, 236, 237
Alarm
Alarm
366, 369, 370
Alarm
Alarm
Alarm
Warning
181, 197, 198,
276
Warning
381, 213, 214,
395
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
Warning
122, 363
Warning
254, 495, 496
Warning
Warning
39
Warning
Warning
Description
Temperature of the TEC#1 (181) is
outside of a band set by parameter 183
around the setpoint (188)
Temperature of the TEC#2 (374) is
outside of a band set by parameter 376
around the setpoint (381)
Cell temperature (192) is outside of a
band set by parameter 194 around the
setpoint (193)
Pressure transducer reading (244) is
outside of a band set by min/max
pressure (236, 237)
For ATEX ONLY: SEAL pressure (366)
is outside of a band set by min/max
pressure (369, 370)
File system error, reported in system
alarm (125)
Internal harware error
TEC temperature (181) drifts out of
factory setting window (197, 198)
defined by the factory default (276)
TEC temperature 2 (381) drifts out of
factory setting window (213, 214)
defined by the factory default (395)
Concentration reading from cell #1 is out
of calibration range
Concentration reading from cell #2 is
out of calibration range
Laser intensity in sample cell is low
Internal battery failure
Laser intensity in reference cell is low
Concentration in reference cell is low
Concentration alarm from band 1
(channel 1)
Concentration alarm from band 2
(channel 1)
Concentration alarm from band 3
(channel 1)
Concentration alarm from band 4
(channel 1)
Concentration alarm from band 1
(channel 2)
Concentration alarm from band 2
(channel 2)
Concentration alarm from band 3
(channel 2)
Concentration alarm from band 4
(channel 2)
Concentration reading is invalid.
Possible causes include: system in line-
lock, system in calibration, etc...
Cell pressure (254) is outside of a band
set by min/max pressure (495, 496)
Error in span calibration
Error in zero calibration
User error detected from the user analog
input
Maintenance
|
6-15