Analog Range Buttons; Test Config Button - Ametek 2850 User Manual

Moisture analyzer
Hide thumbs Also See for 2850:
Table of Contents

Advertisement

Analog Range Button
ANALOG
RANGE
To access the Analog Range Menu, press the
item (variable) is highlighted and press
Range Select Out1
View Range Limits
Alter Range Limits
NOTE
Range Limits to be viewed or altered are:
Output_1_Low
Output_1_Hi
Output_2
Offset_1_Low
Offset_1_Hi
Offset_2
Test / Config Button
TEST
CONFIG
With the exception of alarm levels and range limits, all system configuration parameters, both critical (system con-
stants and variables, calculation coefficients, configuration flags, etc.) and non-critical (time of day, date, etc.), are
accessed through the
test
will be required immediately. If the Level 1 access code has been entered, a higher level access code will be needed
to select Configure Menu. The menu accessed when Configure Menu is chosen will be different depending on
which access code, Level 2 or Level 3, has been entered.
4-8 2850 Moisture Analyzer
analog range
.
enter
View and/or alter current analog range (low or high) on analog output 1. When this menu
item is selected, the third line of the display will say either "Range is now Low" or "Range
is now Hi". Press
cancel
press
to toggle the range. The current range message will be shown momentarily
enter
and then the display will return to the Range Menu.
To view range limits, highlight the limit to be viewed from the following list and press
to select. Press
enter
The Level 1 or higher access code is required for this function. Highlight the limit to be
altered from the following list and press
the "New Value?" prompt is displayed. Press
list; press
to quit.
cancel
The engineering units associated with the analog range limits are determined by the pa-
rameter linked to the output. If moisture concentration in ppm is on the output, range
limits are also defined in ppm; if cell temperature in °C is on output, range limits are
also in °C and so on.
20 mA value of channel 1 low output range
20 mA value of channel 1 high output range
20 mA value of channel 2 output range
4 mA value of channel 1 low output range
4 mA value of channel 1 high output range
4 mA value of channel 2 output range
/
button. When the
config
button. Scroll through the menu until the desired menu
to return to Analog Range Menu without changing the range;
to return to limit list.
cancel
to select. Enter the new limit value when
enter
enter
/
button is pressed, an access code (level 1, 2, or 3)
test
config
to save change and return to the limit