Analog Range

The Analog Range key defines your 4 to 20 mA output setting. Use this setting
to scale the 4 to 20 mA output. The 4 to 20 mA output is proportional to the read-
ings.
Main menu for Analog Range key.

Max Concentration

Enter the reading equivalent to 20 mA.

Min Concentration

Enter the reading equivalent to 4 mA.
Example:
To scale the analog output to represent a 0 to 100 PPMV range enter the follow-
ing settings:
Max Concentration
Min Concentration
4-8 | Model 5100 HD Analyzer
ANALOG
MAX CONCENTRATION
RANGE
MIN CONCENTRATION
100
0
*
*