2-4 | Model 5100 HD Analyzer
Sample Temperature
Ambient temperature
Storage temperature
Ambient humidity
Physical
Enclosure material
Maximum altitude
Pollution Degree
Ingress protection (IP)
Sample flow rate
Sample gas
Sample wetted materials
Electrical Classifications
-20C to +150C (-4 F to +302 F) Gas-phase samples
only
-20
C to +50
C (-4
F to +122
o
o
o
-25 C to +80 C (-13 F to +176 F)
0-90%, non-condensing
880 mm x 674 mm x 302 mm (34.6" x 26.5" x 11.9"),
60 kgs (132 lbs)
300 series stainless steel
2000 meters
2
IP65 and NEMA Type 4X
1-10 lpm (2-20 SCFH)
Natural gas, Hydrogen, Carbon Dioxide, Propane,
Oxygen, and others pending application review
Stainless steel, SiO2 glass, Gold, Teflon, EPDM
rubber
This analyzer will be certified with regard to
electrical safety for Pollution Degree 2 (normally
non-conductive environmental pollution occurs
with occasional condensation) and Installation
Category II (local level transients, less than
those found at power distribution levels). Mains
supply voltage fluctuations are not to exceed 10
percent of the nominal supply voltage. Other ap-
provals will include acceptance for use in
NEC/CEC Class I Division 2 Groups A, B, C, D T3
and world-wide group II Zone 2 IIC T3 classified
hazardous areas.