Controller Communications

1-8 | Model 5100 HD Analyzer
All analyzer functions are controlled by microprocessors housed within the
electronics enclosure. These include:
(1) Non-isolated Analog Input (4 to 20 mA or 0-5 Volts)
(1) Isolated discrete DC Input
(1) Isolated 4-20 mA analog output per laser diode (i.e. one output for single
laser analyzer and (2) outputs for dual laser analyzer)
(4) Solenoid Valve Drives
(4) Relay Contacts
(2) RTD Inputs
Moisture Reference Cell
Piezo-Resistive Transducer Input
(3) Solid State Relay Drives (PWM)
Single Isolated Discrete Input
(3) Serial Ports: (1) RS-232, (2) RS-485
100 Mbs Ethernet Port
MQX RTOS and Modbus capability
The Model 5100 HD is configured, and analyzer parameters are set, using either
the analyzer display and keypad, or the Web Interface. The 4-line x 20-character
VFD provides moisture data, system operating parameters, system messages, and
alarm conditions. The 20-button keypad on the front of the unit is used to set or
change operating parameters and system values. There is a serial interface inside
the main electronics enclosure, which is intended for service use only.