Ametek 5100 HD User Manual page 117

Tunable diode laser spectrometer
Hide thumbs Also See for 5100 HD:
Table of Contents

Advertisement

139
rw
uint16
ANAOUT1OR
140
rw
uint16
ANAOUT1F
141
r
float
ANAOUT2
142
rw
float
ANAOUT2O
143
rw
uint16
ANAOUT2P
144
rw
float
ANAOUT2S
145
rw
int16
ANAOUT2T
146
rw
uint16
ANAOUT2OR
147
rw
uint16
ANAOUT2F
148
r
uint16
RELAY1
149
rw
uint16
RELAY1D
150
rw
uint16
RELAY1F
151
rw
uint16
RELAY1S
152
rw
float
RELAY1V
153
r
uint16
RELAY2
154
rw
uint16
RELAY2D
155
rw
uint16
RELAY2F
156
rw
uint16
RELAY2S
157
rw
float
RELAY2V
158
r
uint16
RELAY3
159
rw
uint16
RELAY3D
160
rw
uint16
RELAY3F
161
rw
uint16
RELAY3S
162
rw
float
RELAY3V
163
r
uint16
RELAY4
164
rw
uint16
RELAY4D
165
rw
uint16
RELAY4F
166
rw
uint16
RELAY4S
167
rw
float
RELAY4V
0 = 3.78 - 20.5, 1 = 0-20.5, high or low
values can also be set by fault value, See
Fault option (ANAOUT1F)
133
1
Current Output 1 Over range Enable.
0=disabled, 1=enabled
134
1
Current Output 1 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-20 mode
(ANAOUT1T = 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
135
2
Current Output 2 -
137
2
Current Output 2 Offset -
139
1
Current Output 2 Parameter-
140
2
Current Output 2 Span -
142
1
Current Output 2 Output Type. 0=4-20 mode,
1=0-20 mode.
If overange enable (ANAOUT2OR = 1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low
values can also be set by fault value, See
Fault option (ANAOUT2F) -
143
1
Current Output 2 Over range Enable.
0=disabled, 1=enabled
144
1
Current Output 2 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-20 mode
(ANAOUT2T = 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
145
1
Relay 1. 0=OPEN, 1=CLOSED
146
1
Relay 1 Delay. Holds the relay output for this
set amount of time.
147
1
Relay 1 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
148
1
Relay 1 Source. The source parameter
number used to evaluated the relay function.
149
2
Relay 1 Value. The value used to evaluate
the relay function.
151
1
Relay 2. 0=OPEN, 1=CLOSED
152
1
Relay 2 Delay. Holds the relay output for this
set amount of time.
153
1
Relay 2 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
41
alarm
154
1
Relay 2 Source. The source parameter
number used to evaluated the relay function.
155
2
Relay 2 Value. The value used to evaluate
the relay function.
157
1
Relay 3. 0=OPEN, 1=CLOSED
158
1
Relay 3 Delay. Holds the relay output for this
set amount of time.
159
1
Relay 3 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
160
1
Relay 3 Source. The source parameter
number used to evaluated the relay function.
161
2
Relay 3 Value. The value used to evaluate
the relay function.
163
1
Relay 4. 0=OPEN, 1=CLOSED
164
1
Relay 4 Delay. Holds the relay output for this
set amount of time.
165
1
Relay 4 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
166
1
Relay 4 Source. The source parameter
number used to evaluated the relay function.
167
2
Relay 4 Value. The value used to evaluate
the relay function.
|
Maintenance
6-17