6-8 | Model 5100 HD Analyzer
220
rw
char[20]
ALYTE2
float
ALYTE2CON
float
TECTEMP
int16
TECAT
float
TECDEVT
float
TECDT
float
TECHDC
float
TECIT
float
TECPB
float
TECSP
float
TECST
float
TECHYS
float
ELECTEMP
float
CELLTEMP
float
CELLSP
float
CELLDEVT
float
NORMTEMP
float
NORMPRES
float
MAXTEMP
float
MINTEMP
uint
LOGSAVEINT
char[10]
CONCUNIT2
uint
ALYTE1RES
uint
ALYTE2RES
boolean
RELAY1NC
boolean
RELAY2NC
boolean
RELAY3NC
boolean
RELAY4NC
uint16
MENUTO
uint16
PWDTO
uint
SYSWARN1
uint
SYSWARN2
uint
SYSSTWARN
uint16
PTSEED
float
MAXTEMP_2
float
MINTEMP_2
float
REFCONC_2
int16
STATICPT
float
STCELLT
float
STCELLP
int16
ANAOUT1H
int16
ANAOUT2H
198
10
Analyte 2 Name
208
2
Analyte 2 Concentration
210
2
TEC Temperature Value
212
1
Heater Actuator Type. 0=Reverse,
1=Forward.
213
2
Heater Deviation Tolerance. Ignored if
0. Otherwise high or low limit alarm
are set when value is out bounds of 1
deviation
215
2
Heater Deriviative Time
217
2
Heater Duty Cycle
219
2
Heater Integral Time
221
2
Heater Proportional Band
223
2
Heater Set Point
225
2
TEC Sample Time
227
2
TEC hysterisi
229
2
Electronics Temperature
231
2
Cell RTD1 Filtered value
233
2
Cell Set Point
235
2
Cell Deviation Tolerance. Ignored if 0.
Otherwise high or low limit alarm are
set when value is out bounds of 1
deviation
237
2
Calibration temperature for
concentration compensation.
239
2
Calibration pressure for concentration
compensation
241
2
Max TEC temperature
243
2
Min TEC temperature
245
2
Interval in minutes to save the log
data
247
5
Concentration Unit of Measure for
analyte 2
252
2
Decimal places to use for displaying
analyte 1
254
2
Decimal places to use for displaying
analyte 2
256
1
Normally closed indicator for relay 1.
31
1=Normally closed, 0=Normally open
257
1
Normally closed indicator for relay 2.
1=Normally closed, 0=Normally open
258
1
Normally closed indicator for relay 3.
1=Normally closed, 0=Normally open
259
1
Normally closed indicator for relay 4.
1=Normally closed, 0=Normally open
260
1
Menu timeout. The time that the menu
will navigate to the homescreen after
no activity. 0 is disabled
261
1
Password timeout. The time that a
user logged into the front display
menu will be logged out after no
activity. 0 is disabled
262
2
Temporary cache to miror warnings in
SYSALARM1
264
2
Temporary cache to miror warnings in
SYSALARM2
266
2
system statuc warning
268
1
The seed value for all IIR filters.
269
2
Max TEC temperature 2
271
2
Min TEC temperature 2
273
2
Concentration of Reference call
275
1
Pressure (STCELLP) and temp
(STCELLT) static value usage.
0=neither, 1=use static pressure,
2=use static temperature, 3= use both
276
2
Static cell temperature. Use by
setting STATICPT to 2 or 3
278
2
Static cell pressure. Use by setting
STATICPT to 1 or 3
280
1
Analog output 1, 0 - follow source
during cal, 1 - hold output during cal
281
1
Analog output 2, 0 - follow source