Power and Environmental Conditions

Transient Withstand Capability

C A U T I O N
C A U T I O N
358

Maximum input voltage for analog inputs

CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with N2863A 10:1 probe: CAT I 600 V, CAT II 300 V (DC + peak AC)
with 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
with 50
input: 5 Vrms
Do not exceed 5 Vrms in 50
protection is enabled in 50
if greater than 5 Vrms is detected. However, the input could still be
damaged, depending on the time constant of the signal. The 50
protection mode only functions when the oscilloscope is powered on.
mode on the 2-channel models. Input
mode, and the 50
load will disconnect