4
Triggering the Oscilloscope
150
Error Frame - The oscilloscope will trigger on CAN active
error frames.
All Errors - The oscilloscope will trigger when any form
error or active error is encountered.
Acknowledge Error - The oscilloscope will trigger when the
acknowledge bit is recessive (high).
Overload Frame - The oscilloscope will trigger on CAN
overload frames.
3 If necessary, press the up-arrow softkey to return to the CAN
Trigger menu. Press the Signals softkey to enter the CAN
Signals menu.
Signals
Baud Rate
source
4 Press the Source softkey to select the channel connected to
the CAN signal line.
As you repeatedly press the Source softkey (or rotate the
Entry knob), the CAN label for the source channel is
automatically set and the channel you select is shown in the
upper-right corner of the display next to "CAN".
Adjust the trigger level for the selected analog channel by
turning the Trigger Level knob. Press the D15 Thru D0 key and
select Thresholds to set the threshold level for digital
channels. The value of the trigger level or digital threshold is
displayed in the upper-right corner of the display.
CAN trigger
Sample
Point
6000 Series Oscilloscope User's Guide
Currently selected
Trigger level
signal source
or threshold
Return to
previous menu